Medtronic MiniMed Paradigm Insulin pump, model number… recallZ-1587-2020 · Medtronic MiniMed · initiated Jun 27, 2019
Class II Open, Classified All recalls posted in March 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic MiniMed or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.
FDA root cause category: Device Design
What the company is doing
On June 27, 2019, Medtronic issued an Urgent Field Safety Notification letter via mail and email to its consignees. Medtronic made the decision to renotify customers, and on January 4, 2023, Medtronic issued an Urgent Medical Device Correction - Update letter. The communication asks consignees to take the following actions: .To find the software version for the MiniMed Paradigm pumps, go to the STATUS screen: . To open the STATUS screen, press ESC until the STATUS screen appears. . To view more text on the STATUS screen, press the up or down arrow to scroll and view all the information. . To exit the STATUS screen, press ESC until the STATUS screen disappears. ACTIONS REQUIRED: 1. Review and take the following cybersecurity precautions listed below. 2. Please complete and sign the customer confirmation form attached as indicated and return by mail using the attached prepaid envelope. You can also email this form to our Field Corrective Action Department at [email protected] or send via fax to 1-888-844-8084 to the attention of FCA Department.
Recall details
- Recall number
- Z-1587-2020
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jun 27, 2019
- Posted by the FDA
- Mar 26, 2020
- Recalling firm
- Medtronic MiniMed
- Quantity
- 88,243
- Product code
- OYC
- FDA submissions
- P980022
- Distribution
- US nationwide distribution including Puerto Rico.
Product description
Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
Code information (models, lots, serial numbers)
All software versions
Other devices under the same FDA submission
277 devices cleared under P980022 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Paradigm® REAL-Time Revel™ MMT-723NASMedtronic Minimed, Inc. · Sensor, glucose, invasive15 deaths3,413 reports
- Paradigm® REAL-Time Revel™ MMT-723LNASMedtronic Minimed, Inc. · Sensor, glucose, invasive1 recall12 deaths2,995 reports
- Paradigm® REAL-Time Revel™ MMT-723LNASMedtronic Minimed, Inc. · Sensor, glucose, invasive1 recall18 deaths2,909 reports
- Paradigm® REAL-Time Revel™ MMT-723LNASMedtronic Minimed, Inc. · Sensor, glucose, invasive1 recall23 deaths2,570 reports
- Paradigm® REAL-Time Revel™ MMT-723RNASMedtronic Minimed, Inc. · Sensor, glucose, invasive18 deaths1,970 reports
- Paradigm® REAL-Time Revel™ MMT-723NAPMedtronic Minimed, Inc. · Sensor, glucose, invasive7 deaths1,564 reports
- Paradigm® REAL-Time Revel™ MMT-723NALMedtronic Minimed, Inc. · Sensor, glucose, invasive9 deaths1,504 reports
- Paradigm® REAL-Time Revel™ MMT-723NABMedtronic Minimed, Inc. · Sensor, glucose, invasive8 deaths1,492 reports
- Paradigm® REAL-Time Revel™ MMT-723LNALMedtronic Minimed, Inc. · Sensor, glucose, invasive9 deaths1,271 reports
- Paradigm® REAL-Time Revel™ MMT-523NASMedtronic Minimed, Inc. · Sensor, glucose, invasive9 deaths1,259 reports
- Paradigm® REAL-Time Revel™ MMT-723LNAPMedtronic Minimed, Inc. · Sensor, glucose, invasive7 deaths1,256 reports
- Paradigm® REAL-Time Revel™ MMT-723LNABMedtronic Minimed, Inc. · Sensor, glucose, invasive10 deaths1,243 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 83433).
- Z-1588-2020 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722
- Z-1589-2020 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-522K
- Z-1590-2020 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722K
- Z-1591-2020 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523
- Z-1592-2020 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523K
- Z-1593-2020 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723K
- Z-1594-2020 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page