Monaco recallZ-2409-2019 · Elekta, Inc. · initiated Aug 23, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions.
FDA root cause category: Device Design
What the company is doing
Elekta Inc. notified customers on about 08/23/2019 via "IMPORTANT FIELD SAFETY NOTIFICATION" letter. The letter informed customers that the Monaco is displaying some incorrect shift information on the Scan and Setup Reference Report and is DICOM exporting incorrect shift information. Instructions included that all shifts should be verified before treatment, that a software fix is planned. Customer were also instructed to post the notice in a place accessible to all users, advise personnel working with the device of the issue, and to complete and return the Acknowledgement Form provided.
Recall details
- Recall number
- Z-2409-2019
- Status
- Terminated
- Initiated by the company
- Aug 23, 2019
- Recalling firm
- Elekta, Inc.
- Quantity
- 49 units
- Product code
- MUJ
- FDA submissions
- K190178
- Distribution
- Distribution within the US to Ohio and Indiana. International distribution to Germany, Greece, Sweden.
Product description
Monaco Radiation Treatment Planning System (RTP) System
Code information (models, lots, serial numbers)
Software Build: 5.50.00 & 5.51.00; UDI GTIN: (01)00858164002244(10) 5.50.00, (01)00858164002268(10) 5.51.00
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.