Monaco recallZ-2409-2019 · Elekta, Inc. · initiated Aug 23, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions.

FDA root cause category: Device Design

What the company is doing

Elekta Inc. notified customers on about 08/23/2019 via "IMPORTANT FIELD SAFETY NOTIFICATION" letter. The letter informed customers that the Monaco is displaying some incorrect shift information on the Scan and Setup Reference Report and is DICOM exporting incorrect shift information. Instructions included that all shifts should be verified before treatment, that a software fix is planned. Customer were also instructed to post the notice in a place accessible to all users, advise personnel working with the device of the issue, and to complete and return the Acknowledgement Form provided.

Recall details

Recall number
Z-2409-2019
Status
Terminated
Initiated by the company
Aug 23, 2019
Recalling firm
Elekta, Inc.
Quantity
49 units
Product code
MUJ
FDA submissions
K190178
Distribution
Distribution within the US to Ohio and Indiana. International distribution to Germany, Greece, Sweden.

Product description

Monaco Radiation Treatment Planning System (RTP) System

Code information (models, lots, serial numbers)

Software Build: 5.50.00 & 5.51.00; UDI GTIN: (01)00858164002244(10) 5.50.00, (01)00858164002268(10) 5.51.00