Skip to content

Monaco (UDI 00858164002244)Recalls & reported problems

Elekta, Inc. · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Nov 22, 2019Open, Classified

    Monaco is using the incorrect energy when optimizing and calculating dose.

    Z-0670-2020 · Root cause: Software Design Change · FDA record ↗

  2. Sep 4, 2019Open, Classified

    It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

    Z-2561-2019 · Root cause: Software design · FDA record ↗

  3. Aug 23, 2019Terminated

    Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions.

    Z-2409-2019 · Root cause: Device Design · FDA record ↗

Other System, Planning, Radiation Therapy Treatment devices

All System, Planning, Radiation Therapy Treatment devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.