Monaco recallZ-2561-2019 · Elekta Inc · initiated Sep 4, 2019

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

FDA root cause category: Software design

What the company is doing

The firm notified their consignees of the issue by email on 09/04/2019. The letter explained the problem and the workaround and requested the notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed and that all appropriate personnel working with the product be advised of the content of the letter.

Recall details

Recall number
Z-2561-2019
Status
Open, Classified
Initiated by the company
Sep 4, 2019
Recalling firm
Elekta Inc
Quantity
49 units
Product code
MUJ
FDA submissions
K190178
Distribution
Worldwide Distribution - US Nationwide in the states of WI, OH, NY, OH, IN, IA, TX, and countries of Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, UK

Product description

Monaco RTP System, 5.50/5.51, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review

Code information (models, lots, serial numbers)

UDI: (01)00858164002244(10) 5.50.00  (01)00858164002268(10) 5.51.00

Devices named in this recall