CardioChek Plus Professional Analyzer recallZ-0832-2020 · Polymer Technology Systems, Inc. · initiated Dec 2, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Polymer Technology Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
FDA root cause category: Device Design
What the company is doing
Urgent Medical Device Correction notification letters dated 11/15/19 were sent to customers. Actions to be taken by the Customer/Distributor/User: 1. Immediately check your inventories and apply the sticker/labels as directed. If you are an end user: Please adhere one label to the outside of the battery door compartment of your analyzer and apply the other label to the analyzer user guide. If you are a distributor: Please remove the tape liner on the back of the large sticker leaving both small labels intact for the end user to apply. Apply the large sticker to the top panel of the analyzer box, and press firmly. 2. If you have further distributed this product, immediately identify the users and notify them at once of this product correction. Forward this notice, stickers, and response form to all customers/users who did, or who may have, received, this product. 3. Once you have taken the appropriate action, please fill out the attached response form and return it to PTS Diagnostics via the information below. It is important that you complete and return the response form, even if you do not have affected product on hand. 4. Please contact PTS for additional stickers and a PDF version of this letter and response form, if needed. Thank you for your prompt action, and we apologize for any inconvenience this issue may have caused you and your facility. If you have any questions, please contact Customer Service, Monday through Friday, 8:00 AM to 5:00 PM, Eastern Time at +1 319-870-5610 or 877-870-5610 (U.S. toll-free), or via email at [email protected].
Recall details
- Recall number
- Z-0832-2020
- Status
- Terminated
- Initiated by the company
- Dec 2, 2019
- Recalling firm
- Polymer Technology Systems, Inc.
- Quantity
- 44,448 total
- Product code
- NBW
- FDA submissions
- K140068
- Distribution
- The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.
Product description
PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700
Code information (models, lots, serial numbers)
All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00381932700016
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CardioChek® Plus Professional Analyzer 2700Polymer Technology Systems, Inc. · LDL & VLDL PRECIPITATION, HDL1 recall114 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K140068 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CardioChek 5700411Henry Schein, Inc. · ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL1 recall5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 84274).
- Z-0833-2020 · TerminatedPTS Diagnostics CardioChek Plus Professional Analyzer, Refurbished analyzer, without box or carrying case, Reference Numbers 2702
- Z-0834-2020 · TerminatedHenry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036
- Z-0835-2020 · TerminatedHenry Schein CardioChek Plus Professional Analyzer, New Analyzer, boxed, with carrying case, Reference Number 4060 HS, 570-0411
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.