Monaco recallZ-0670-2020 · Elekta Inc · initiated Nov 22, 2019

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Monaco is using the incorrect energy when optimizing and calculating dose.

FDA root cause category: Software Design Change

What the company is doing

An Important Field Safety Notification letter is being emailed to affected customers beginning 11/22/2019. The letter identifies affected product, provides a description of the problem, and recommends user not use multiple energies within the same plan until a software fix has been applied to the device. A response form was asked to be returned. Affected customers are advised to contact local Elekta service teams for questions.

Recall details

Recall number
Z-0670-2020
Status
Open, Classified
Initiated by the company
Nov 22, 2019
Recalling firm
Elekta Inc
Quantity
76 (7 US, 69 OUS)
Product code
MUJ
FDA submissions
K183037, K190178
Distribution
IA, IN, NY, OH, TX, WI. Internationally distributed to: Australia, Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, Switzerland, United Kingdom.

Product description

Monaco Radiation Treatment Planning System (RTP) System

Code information (models, lots, serial numbers)

Software Builds: 5.40.00, 5.50.00 and 5.51.00. UDI GTIN: (01)00858164002190(10) 5.40.00, (01)00858164002244(10) 5.50.00, and (01)00858164002268(10) 5.51.00

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates