CareFusion Alaris System, Alaris PC unit models 8000 and… recallZ-1359-2020 · CareFusion 303, Inc. · initiated Feb 4, 2020
Class I Open, Classified All recalls posted in March 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders
FDA root cause category: Device Design
What the company is doing
On 02/04/20, Urgent Medical Device Recall Notifications were mailed and posted on the firm's website: www.bd.com/alaris-system-software-recall Two notifications were posted: 1) Software versions 9.33 and prior, and 2) software version 12.1.0. On 02/05/20. Recall letters can be found at https://www.bd.com/en-us/recall-notifications/detail-mms-20-1953 Firm will contact all customers to initiate the scheduling process for the software update. Before the update customers were asked to so the following: 1) Software Errors related to System Error 255-XX-XXX: If the error occurs while administering a critical medication(s), continue the infusion while you expedite a replacement pump if one is readily available, or restart and reprogram the PC unit. 2) KVO Rate Not Available When Using Delay Options Programming: For software versions 9.19 and prior: Set a Callback alert of "After" or "Before and After." For all software versions: Do not use Delay Options when a KVO rate is required. 3) Low Battery Alarm Failure: Whenever possible, keep the PC unit plugged into AC power. 4) Keep Vein Open (KVO) / End of Infusion Alarms Priority: Check and increase the audio volume. 5) Use Errors related to Custom Concentration Programming: Only use custom concentration when the medication label does not match any of the drug concentration selections on the programming screen. Customers can call the following phone numbers for assistance: Customer Advocacy at 888-812-3266, Clinical and Pharmacy Support Center at 858-617-1316, Recall Support Center 888-562-6018, and Technical Support 888-812-3229. On 07/29/21, Follow up Medical Device Recall Notifications were mailed to customers informing them that a new software version, v12.1.2., is available to remediate this recalls issues. Complete the acknowledgement form and contact the Recall Support Center to schedule remediation.
Recall details
- Recall number
- Z-1359-2020
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Feb 4, 2020
- Posted by the FDA
- Mar 5, 2020
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 870070
- Product code
- FRN
- FDA submissions
- K133532
- Distribution
- Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN, AL, NH, OK, DC, NJ, MO, SD, UT, NM, MD, VT, MP, ND, AS; and countries of: Canada, Australia, New Zealand, the Middle East, South Africa, Taiwan, Southeast Asia and the European Union.
Product description
CareFusion Alaris System, Alaris PC unit models 8000 and 8015 (UDI: 10885403801518)
Code information (models, lots, serial numbers)
Model 8000, software versions 9.5 and prior; Model 8015, software versions 12.1.0, and 9.33 and prior
Other devices under the same FDA submission
4 devices cleared under K133532 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion1 recall2 deaths5,636 reports
- Alaris 8975Carefusion 303, Inc. · Infusion safety management software169 reports
- Alaris 9601Carefusion 303, Inc. · Pump, infusion1 recall0 reports
- Alaris™ Syringe Module, Model 8110 8100Carefusion 303, Inc. · Pump, infusion14 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 84579).
- Z-1360-2020 · Open, ClassifiedCareFusion Alaris Pump Module, Model 8100
- Z-1361-2020 · Open, ClassifiedCareFusion Alaris Syringe Module, Model 8110
- Z-1362-2020 · Open, ClassifiedCareFusion Alaris PCA Module, Model 8120
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page