CareFusion Alaris Syringe Module, Model 8110 recallZ-1361-2020 · CareFusion 303, Inc. · initiated Feb 4, 2020
Class I Open, Classified All recalls posted in March 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders
FDA root cause category: Device Design
What the company is doing
On 02/04/20, Urgent Medical Device Recall Notifications were mailed and posted on the firm's website: www.bd.com/alaris-system-software-recall Two notifications were posted: 1) Software versions 9.33 and prior, and 2) software version 12.1.0. On 02/05/20. Recall letters can be found at https://www.bd.com/en-us/recall-notifications/detail-mms-20-1953 Firm will contact all customers to initiate the scheduling process for the software update. Before the update customers were asked to so the following: 1) Software Errors related to System Error 255-XX-XXX: If the error occurs while administering a critical medication(s), continue the infusion while you expedite a replacement pump if one is readily available, or restart and reprogram the PC unit. 2) KVO Rate Not Available When Using Delay Options Programming: For software versions 9.19 and prior: Set a Callback alert of "After" or "Before and After." For all software versions: Do not use Delay Options when a KVO rate is required. 3) Low Battery Alarm Failure: Whenever possible, keep the PC unit plugged into AC power. 4) Keep Vein Open (KVO) / End of Infusion Alarms Priority: Check and increase the audio volume. 5) Use Errors related to Custom Concentration Programming: Only use custom concentration when the medication label does not match any of the drug concentration selections on the programming screen. Customers can call the following phone numbers for assistance: Customer Advocacy at 888-812-3266, Clinical and Pharmacy Support Center at 858-617-1316, Recall Support Center 888-562-6018, and Technical Support 888-812-3229. On 07/29/21, Follow up Medical Device Recall Notifications were mailed to customers informing them that a new software version, v12.1.2., is available to remediate this recalls issues. Complete the acknowledgement form and contact the Recall Support Center to schedule remediation.
Recall details
- Recall number
- Z-1361-2020
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Feb 4, 2020
- Posted by the FDA
- Mar 5, 2020
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 114835
- Product code
- FRN
- FDA submissions
- K051641, K012383
- Distribution
- Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN, AL, NH, OK, DC, NJ, MO, SD, UT, NM, MD, VT, MP, ND, AS; and countries of: Canada, Australia, New Zealand, the Middle East, South Africa, Taiwan, Southeast Asia and the European Union.
Product description
CareFusion Alaris Syringe Module, Model 8110
Code information (models, lots, serial numbers)
Software versions 12.1.0, and 9.33 and prior
Other devices under the same FDA submission
1 device cleared under K051641, K012383 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,093 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 84579).
- Z-1359-2020 · Open, ClassifiedCareFusion Alaris System, Alaris PC unit models 8000 and 8015 (UDI: 10885403801518)
- Z-1360-2020 · Open, ClassifiedCareFusion Alaris Pump Module, Model 8100
- Z-1362-2020 · Open, ClassifiedCareFusion Alaris PCA Module, Model 8120
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page