Skip to content

Probe, 9 FR AUTOLITH recallZ-1883-2020 · Northgate Technologies, Inc. · initiated Mar 10, 2020

Class II Terminated All recalls posted in March 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Northgate Technologies, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

As part of an investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Northgate Technologies, Inc. notified customers on 03/11/2020 via "Urgent: medical Device Recall Notification" email. The recall notice identified the affected and requested the customers to cease distribution and destroy affected product. The customers were asked to complete the "Recall Acknowledgment Form".

Recall details

Recall number
Z-1883-2020
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 10, 2020
Posted by the FDA
Mar 24, 2020
Recalling firm
Northgate Technologies, Inc.
Quantity
298 boxes. 3 units per box
Product code
FFK
FDA submissions
K130368
Distribution
US Nationwide distribution. No governmental.

Product description

Probe, 9 FR AUTOLITH; NTI Catalog Number 9-900-54 - Product Usage: Probes are used for the fragmentation of urinary/renal and biliary calculi.

Code information (models, lots, serial numbers)

NTI Catalog Number 9-900-54.

Other devices under the same FDA submission

3 devices cleared under K130368 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 85204).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page