Skip to content

MOSAIQ, oncology information system - Product Usage: MOSAIQ… recallZ-1597-2020 · Elekta Impac Software · initiated Sep 24, 2019

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta Impac Software or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The user may inadvertently enter Metric values into Height and Weight fields labeled with US Standard Units.

FDA root cause category: Software design

What the company is doing

The firm began notifying their consignee on 09/24/2019 by email. The notice explained the issue and provided a workaround for the consignee to use until the updated software is ready for release.

Recall details

Recall number
Z-1597-2020
Status
Open, Classified
Initiated by the company
Sep 24, 2019
Posted by the FDA
Mar 26, 2020
Recalling firm
Elekta Impac Software
Quantity
3732 units
Product code
MUJ
FDA submissions
K141572
Distribution
Worldwide distribution.

Product description

MOSAIQ, oncology information system - Product Usage: MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery.

Code information (models, lots, serial numbers)

Software Versions (UDI/GTIN): MOSAIQ 2.00 (00858164002305); MOSAIQ 2.20 (00858164002312); MOSAIQ 2.30 (00858164002329); MOSAIQ 2.40 (00858164002077); MOSAIQ 2.41 (00858164002084); MOSAIQ 2.60 -2.64 (00858164002091); MOSAIQ 2.50 (00858164002107); MOSAIQ 2.65 (00858164002220); MOSAIQ 2.70 (00858164002037); MOSAIQ 2.80 (00858164002251); MOSAIQ 2.81 (00858164002282)

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates