Skip to content

StageOne Select recallZ-1640-2020 · Biomet, Inc. · initiated Feb 26, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.

FDA root cause category: Process control

What the company is doing

On 2/26/2020, Zimmer Biomet sent an "Urgent Medical Device Recall" letter via FEDEX to direct sales distributors, hospitals, and surgeons. In addition, an email was sent to all distributors alerting them of the recall. Customers were instructed to quarantine the product and a Zimmer Biomet representative would remove the affected product from the facility. They asked customers to complete and return the "Certificate of Acknowledgement" form indicating whether they had any of the affected lots.

Recall details

Recall number
Z-1640-2020
Status
Terminated
Initiated by the company
Feb 26, 2020
Posted by the FDA
Apr 3, 2020
Recalling firm
Biomet, Inc.
Quantity
230,001
Product code
KWY
FDA submissions
K080979, K161166
Distribution
Worldwide distribution. US distribution to the following states: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, West Virginia, Wyoming, and the District of Columbia.

Product description

BIOMET StageOne Hip Cement Spacer Mold with Insert, 56 MM, Silicone, Sterile, Item 431183.

Code information (models, lots, serial numbers)

Show the full code information (8k characters)

***Lot Numbers*** 082090, 651800, 982380, 809020, 320900, 623230, 623240, 630160, 636040, 823140, 234220, 370430, 556610, 047420, 846270, 270160, 374980, 382930, 508920, 419300, 189910, 404850, 844540, 474080, 009850, 467650, 297630, 280110, 913710, 913720, 916350, 039560, 364320, 858200, 557640, 990120, 328020, 102150, 486050, 903730, 295200, 906770, 906940, 907000, 109280, 952910, 627830, 496520, 972210, 038630, 112390, 496530, 627840, 292590, 401300, 441840, 362300, 683630, 966950, 074190, 184490, 829820, 949770, 591260, 988950, 322780, 757440, 827110, 827090, 229610, 404850R, 892790, 673830, 229600, 827060, 757460, 892730, 892750, 757430, 757450, 827080, 628570, 808650, 628560, 808640, 884540, 884530, 757420, 757490, 765160, 456220, 456250, 456260, 456280, 456300, 456350, 456190, 456370, 483850, 483950, 638590, 638640, 478960, 478970, 478990, 479020, 479040, 483830, 483860, 483880, 483900, 483920, 483940, 483960, 483970, 483990, 569060, 569080, 569090, 569110, 569130, 569140, 569160, 569210, 569240, 570120, 570130, 638550, 638560, 638580, 638600, 638620, 638630, 638650, 638670, 638680, 638690, 638700, 638710, 765090, 765110, 765120, 765150, 765180, 765190, 765210, 765230, 765240, 765260, 765270, 569100, 988390, 483910, 638540, 979260, 981280, 989340, 988640 989850, ***UDI Numbers*** (01)00880304521254(17)170728(10)082090, (01)00880304521254(17)171228(10)651800, (01)00880304521254(17)171221(10)982380, (01)00880304521254(17)171226(10)809020, (01)00880304521254(17)180118(10)320900, (01)00880304521254(17)180213(10)623230, (01)00880304521254(17)180215(10)623240, (01)00880304521254(17)180221(10)630160, (01)00880304521254(17)180220(10)636040, (01)00880304521254(17)180307(10)823140, (01)00880304521254(17)180430(10)234220, (01)00880304521254(17)180603(10)370430, (01)00880304521254(17)180620(10)556610, (01)00880304521254(17)180809(10)047420, (01)00880304521254(17)180811(10)846270, (01)00880304521254(17)180903(10)270160, (01)00880304521254(17)180927(10)374980, (01)00880304521254(17)190102(10)382930, (01)00880304521254(17)190113(10)508920, (01)00880304521254(17)190220(10)419300, (01)00880304521254(17)190404(10)189910, (01)00880304521254(17)190416(10)404850, (01)00880304521254(17)190513(10)844540, (01)00880304521254(17)190603(10)474080, (01)00880304521254(17)190820(10)009850, (01)00880304521254(17)190827(10)467650, (01)00880304521254(17)190916(10)297630, (01)00880304521254(17)191008(10)280110, (01)00880304521254(17)191024(10)913710, (01)00880304521254(17)191025(10)913720, (01)00880304521254(17)191104(10)916350, (01)00880304521254(17)191110(10)039560, (01)00880304521254(17)191223(10)364320, (01)00880304521254(17)200302(10)858200, (01)00880304521254(17)200313(10)557640, (01)00880304521254(17)200315(10)990120, (01)00880304521254(17)200324(10)328020, (01)00880304521254(17)200401(10)102150, (01)00880304521254(17)200416(10)486050, (01)00880304521254(17)200527(10)903730, (01)00880304521254(17)200707(10)295200, (01)00880304521254(17)200824(10)906770, (01)00880304521254(17)200825(10)906940, (01)00880304521254(17)200821(10)907000, (01)00880304521254(17)200915(10)109280, (01)00880304521254(17)200924(10)952910, (01)00880304521254(17)201117(10)627830, (01)00880304521254(17)201202(10)496520, (01)00880304521254(17)201207(10)972210, (01)00880304521254(17)201215(10)038630, (01)00880304521254(17)201214(10)112390, (01)00880304521254(17)201214(10)496530, (01)00880304521254(17)201215(10)627840, (01)00880304521254(17)201230(10)292590, (01)00880304521254(17)210112(10)401300, (01)00880304521254(17)210118(10)441840, (01)00880304521254(17)210202(10)362300, (01)00880304521254(17)210210(10)362310, (01)00880304521254(17)210216(10)727010, (01)00880304521254(17)210302(10)683630, (01)00880304521254(17)210303(10)966950, (01)00880304521254(17)210318(10)074190, (01)00880304521254(17)210322(10)184490, (01)00880304521254(17)210527(10)829820, (01)00880304521254(17)210601(10)949770, (01)00880304521254(17)210824(10)591260, (01)00880304521254(17)220104(10)988950, (01)00880304521254(17)220429(10)322780, (01)00880304521254(17)220620(10)757440, (01)00880304521254(17)220620(10)827110, (01)00880304521254(17)220630(10)827090, (01)00880304521254(17)220725(10)229610, (01)00880304521254(17)190416(10)404850R, (01)00880304521254(17)221102(10)892790, (01)00880304521254(17)221116(10)673830, (01)00880304521254(17)221117(10)229600, (01)00880304521254(17)221117(10)827060, (01)00880304521254(17)221121(10)757460, (01)00880304521254(17)221129(10)892730, (01)00880304521254(17)221130(10)892750, (01)00880304521254(17)221211(10)757430, (01)00880304521254(17)221215(10)757450, (01)00880304521254(17)221103(10)827080, (01)00880304521254(17)230209(10)628570, (01)00880304521254(17)230209(10)808650, (01)00880304521254(17)230217(10)628560, (01)00880304521254(17)230227(10)808640, (01)00880304521254(17)230308(10)884540, (01)00880304521254(17)230416(10)884530, (01)00880304521254(17)221113(10)757420, (01)00880304521254(17)221113(10)757490, (01)00880304521254(17)240815(10)765160, (01)00880304521254(17)240512(10)456220, (01)00880304521254(17)240513(10)456250, (01)00880304521254(17)240513(10)456260, (01)00880304521254(17)240512(10)456280, (01)00880304521254(17)240512(10)456300, (01)00880304521254(17)240513(10)456350, (01)00880304521254(17)240517(10)456190, (01)00880304521254(17)240517(10)456370, (01)00880304521254(17)240525(10)483850, (01)00880304521254(17)240528(10)483950, (01)00880304521254(17)240525(10)638590, (01)00880304521254(17)240608(10)638640, (01)00880304521254(17)240627(10)478960, (01)00880304521254(17)240628(10)478970, (01)00880304521254(17)240628(10)478990, (01)00880304521254(17)240629(10)479020, (01)00880304521254(17)240614(10)479040, (01)00880304521254(17)240627(10)483830, (01)00880304521254(17)240629(10)483860, (01)00880304521254(17)240622(10)483880, (01)00880304521254(17)240630(10)483900, (01)00880304521254(17)240628(10)483920, (01)00880304521254(17)240709(10)483940, (01)00880304521254(17)240628(10)483960, (01)00880304521254(17)240709(10)483970, (01)00880304521254(17)240709(10)483990, (01)00880304521254(17)240710(10)569060, (01)00880304521254(17)240710(10)569080, (01)00880304521254(17)240710(10)569090, (01)00880304521254(17)240627(10)569110, (01)00880304521254(17)240710(10)569130, (01)00880304521254(17)240617(10)569140, (01)00880304521254(17)240630(10)569160, (01)00880304521254(17)240702(10)569210, (01)00880304521254(17)240710(10)569240, (01)00880304521254(17)240627(10)570120, (01)00880304521254(17)240709(10)570130, (01)00880304521254(17)240611(10)638550, (01)00880304521254(17)240711(10)638560, (01)00880304521254(17)240710(10)638580, (01)00880304521254(17)240610(10)638600, (01)00880304521254(17)240629(10)638620, (01)00880304521254(17)240709(10)638630, (01)00880304521254(17)240708(10)638650, (01)00880304521254(17)240608(10)638670, (01)00880304521254(17)240623(10)638680, (01)00880304521254(17)240609(10)638690, (01)00880304521254(17)240620(10)638700, (01)00880304521254(17)240610(10)638710, (01)00880304521254(17)240710(10)765090, (01)00880304521254(17)240628(10)765110, (01)00880304521254(17)240708(10)765120, (01)00880304521254(17)240710(10)765150, (01)00880304521254(17)240710(10)765180, (01)00880304521254(17)240709(10)765190, (01)00880304521254(17)240627(10)765210, (01)00880304521254(17)240621(10)765230, (01)00880304521254(17)240629(10)765240, (01)00880304521254(17)240628(10)765260, (01)00880304521254(17)240710(10)765270, (01)00880304521254(17)240829(10)569100, (01)00880304521254(17)240828(10)988390, (01)00880304521254(17)240916(10)483910, (01)00880304521254(17)240914(10)638540, (01)00880304521254(17)240925(10)979260, (01)00880304521254(17)240925(10)981280, (01)00880304521254(17)240926(10)989340, (01)00880304521254(17)241009(10)988640, (01)00880304521254(17)241011(10)989850

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

14 devices cleared under K080979, K161166 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 85153).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page