STAGEONE recallZ-1659-2020 · Biomet, Inc. · initiated Feb 26, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
FDA root cause category: Process control
What the company is doing
On 2/26/2020, Zimmer Biomet sent an "Urgent Medical Device Recall" letter via FEDEX to direct sales distributors, hospitals, and surgeons. In addition, an email was sent to all distributors alerting them of the recall. Customers were instructed to quarantine the product and a Zimmer Biomet representative would remove the affected product from the facility. They asked customers to complete and return the "Certificate of Acknowledgement" form indicating whether they had any of the affected lots.
Recall details
- Recall number
- Z-1659-2020
- Status
- Terminated
- Initiated by the company
- Feb 26, 2020
- Posted by the FDA
- Apr 3, 2020
- Recalling firm
- Biomet, Inc.
- Quantity
- 230,001
- Product code
- KWS
- FDA submissions
- K131135, K160071
- Distribution
- Worldwide distribution. US distribution to the following states: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, West Virginia, Wyoming, and the District of Columbia.
Product description
BIOMET StageOne, Shoulder Cement Spacer Mold 10 MM, 50 X 21 X 57 MM, Silicone, Sterile, Item 431410.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
***Lot Numbers*** 860790, 215440, 381650, 555570, 089680, 283390, 509720, 657620, 132830, 372720, 097250, 281740, 409230, 409310, 409330, 196720, 196730, 338940, 452050, 581680, 776930, 386800, 233600, 079220, 204510, 262240, 429580, 595980, 849850, 849760, 041390, 041400, 041410, 041420, 041430, 245100, 245110, 245120, 245130, 779360, 915810, 115430, 152270, 326930, 441890, 665300, 924180, 009460, 169730, 389160, 424210, 723180, 723190, 743960, 199850, 970110, 390630, 732410, 398180, 398230, 398190, 398210, 398220, 789770, 789720, 789810, 877270, 202530, 789770R, 789810R, 202530R, 398190R, 674330, 674370, 809050, 325580, 916160, 916170, 916180, 916200, 219810, 398230R, 789720R, 349550, 349560, 349540, 710640, 710650, 846810, 713920, 847000, 847030, 847050, 847140, 846860, 847160, 399490, 399500, 417030, 417010, 417020, 073050, 455050, 704980, 799220, 838250, 918280, 532640, 532660, 028890, 585150, 525420,***UDI Numbers*** (01)00880304551091(17)141007(10)860790, (01)00880304551091(17)141112(10)215440, (01)00880304551091(17)181130(10)381650, (01)00880304551091(17)181231(10)555570, (01)00880304551091(17)190213(10)089680, (01)00880304551091(17)190225(10)283390, (01)00880304551091(17)190320(10)509720, (01)00880304551091(17)190331(10)657620, (01)00880304551091(17)190505(10)132830, (01)00880304551091(17)190528(10)372720, (01)00880304551091(17)190731(10)097250, (01)00880304551091(17)190818(10)281740, (01)00880304551091(17)190826(10)409230, (01)00880304551091(17)190826(10)409310, (01)00880304551091(17)190826(10)409330, (01)00880304551091(17)191103(10)196720, (01)00880304551091(17)191103(10)196730, (01)00880304551091(17)191111(10)338940, (01)00880304551091(17)191119(10)452050, (01)00880304551091(17)191209(10)581680, (01)00880304551091(17)191223(10)776930, (01)00880304551091(17)200314(10)386800, (01)00880304551091(17)200330(10)233600, (01)00880304551091(17)200512(10)079220, (01)00880304551091(17)200521(10)204510, (01)00880304551091(17)200522(10)262240, (01)00880304551091(17)200610(10)429580, (01)00880304551091(17)200706(10)595980, (01)00880304551091(17)200725(10)849850, (01)00880304551091(17)200727(10)849760, (01)00880304551091(17)201111(10)041390, (01)00880304551091(17)201111(10)041400, (01)00880304551091(17)201111(10)041410, (01)00880304551091(17)201110(10)041420, (01)00880304551091(17)201111(10)041430, (01)00880304551091(17)201130(10)245100, (01)00880304551091(17)201130(10)245110, (01)00880304551091(17)201130(10)245120, (01)00880304551091(17)201130(10)245130, (01)00880304551091(17)210119(10)779360, (01)00880304551091(17)210201(10)915810, (01)00880304551091(17)210217(10)115430, (01)00880304551091(17)210224(10)152270, (01)00880304551091(17)210307(10)326930, (01)00880304551091(17)210321(10)441890, (01)00880304551091(17)210407(10)665300, (01)00880304551091(17)210425(10)924180, (01)00880304551091(17)210503(10)009460, (01)00880304551091(17)210517(10)169730, (01)00880304551091(17)210603(10)389160, (01)00880304551091(17)210606(10)424210, (01)00880304551091(17)210705(10)723180, (01)00880304551091(17)210705(10)723190, (01)00880304551091(17)210711(10)743960, (01)00880304551091(17)210808(10)199850, (01)00880304551091(17)210803(10)970110, (01)00880304551091(17)210831(10)390630, (01)00880304551091(17)210929(10)732410, (01)00880304551091(17)220518(10)398180, (01)00880304551091(17)220523(10)398230, (01)00880304551091(17)220523(10)398190, (01)00880304551091(17)220603(10)398210, (01)00880304551091(17)220603(10)398220, (01)00880304551091(17)220601(10)789770, (01)00880304551091(17)220612(10)789720, (01)00880304551091(17)220611(10)789810, (01)00880304551091(17)220608(10)877270, (01)00880304551091(17)220701(10)202530, (01)00880304551091(17)220601(10)789770R, (01)00880304551091(17)220611(10)789810R, (01)00880304551091(17)220701(10)202530R, (01)00880304551091(17)220523(10)398190R, (01)00880304551091(17)221018(10)674330, (01)00880304551091(17)221018(10)674370, (01)00880304551091(17)221106(10)809050, (01)00880304551091(17)221211(10)325580, (01)00880304551091(17)230205(10)916160, (01)00880304551091(17)230204(10)916170, (01)00880304551091(17)230205(10)916180, (01)00880304551091(17)230206(10)916200, (01)00880304551091(17)230330(10)219810, (01)00880304551091(17)220523(10)398230R, (01)00880304551091(17)220612(10)789720R, (01)00880304551091(17)240203(10)349550, (01)00880304551091(17)240205(10)349560, (01)00880304551091(17)240210(10)349540, (01)00880304551091(17)240306(10)710640, (01)00880304551091(17)240228(10)710650, (01)00880304551091(17)240307(10)846810, (01)00880304551091(17)240308(10)713920, (01)00880304551091(17)240307(10)847000, (01)00880304551091(17)240307(10)847030, (01)00880304551091(17)240313(10)847050, (01)00880304551091(17)240309(10)847140, (01)00880304551091(17)240314(10)846860, (01)00880304551091(17)240326(10)847160, (01)00880304551091(17)240416(10)399490, (01)00880304551091(17)240416(10)399500, (01)00880304551091(17)240501(10)417030, (01)00880304551091(17)240502(10)417010, (01)00880304551091(17)240502(10)417020, (01)00880304551091(17)240621(10)073050, (01)00880304551091(17)240628(10)455050, (01)00880304551091(17)240802(10)704980, (01)00880304551091(17)240802(10)799220, (01)00880304551091(17)240801(10)838250, (01)00880304551091(17)240731(10)918280, (01)00880304551091(17)240912(10)532640, (01)00880304551091(17)240912(10)532660, (01)00880304551091(17)241008(10)028890, (01)00880304551091(17)241110(10)585150, (01)00880304551091(17)241109(10)525420, (01)00880304551091(17)220603(10)398210R 398210R.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- STAGEONE 431410Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K131135, K160071 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- STAGEONE 431412Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 recall0 reports
- STAGEONE 431408Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall0 reports
- STAGEONE 431406Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 recall0 reports
- STAGEONE 431414Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85153).
- Z-1638-2020 · TerminatedBIOMET StageOne Hip Cement Spacer Mold with Insert, 48 MM, Silicone, Sterile, Item 431181.
- Z-1639-2020 · TerminatedBIOMET StageOne Hip Cement Spacer Mold with Insert, 52 MM, Silicone, Sterile, Item 431182.
- Z-1640-2020 · TerminatedBIOMET StageOne Hip Cement Spacer Mold with Insert, 56 MM, Silicone, Sterile, Item 431183.
- Z-1641-2020 · TerminatedBIOMET StageOne Hip Cement Spacer Mold with Insert, 60 MM, Silicone, Sterile, Item 431184.
- Z-1642-2020 · TerminatedBIOMET StageOne Hip Cement Spacer Mold with Insert, 64 MM, Silicone, Sterile, Item 431185.
- Z-1643-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 9 X 125 MM, Silicone, Sterile, Item 431190.
- Z-1644-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 11 X 135 MM, Silicone, Sterile, Item 431191.
- Z-1645-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 13 X 145 MM, Silicone, Sterile, Item 431192.
- Z-1646-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 15 X 155 MM, Silicone, Sterile, Item 431193.
- Z-1647-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 17 X 165 MM, Silicone, Sterile, Item 431194.
- Z-1648-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 9 X 200 MM, Silicone, Sterile, Item 431195.
- Z-1649-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 11 X 200 MM, Silicone, Sterile, Item 431196.
- Z-1650-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 13 X 200 MM, Silicone, Sterile, Item 431197.
- Z-1651-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 15 X 200 MM, Silicone, Sterile, Item 431198.
- Z-1652-2020 · TerminatedBIOMET StageOne Select, Hip Stem Cement Spacer Mold w/Reinforcement Cement, 17 X 200 MM, Silicone, Sterile, Item 431199.
- Z-1653-2020 · TerminatedBIOMET StageOne, Hip Cement Spacer Mold w/Reinforcement, 9 X 125 MM, 43 MM, Silicone, Sterile, Item 431207.
- Z-1654-2020 · TerminatedBIOMET StageOne, Hip Cement Spacer Mold w/Reinforcement, 9 X 125 MM, 51 MM, Silicone, Sterile, Item 431209.
- Z-1655-2020 · TerminatedBIOMET StageOne, Hip Cement Spacer Mold w/Reinforcement, 13 X 145 MM, 57 MM, Silicone, Sterile, Item 431213.
- Z-1656-2020 · TerminatedBIOMET StageOne, Hip Cement Spacer Mold w/Reinforcement, 17 X 165 MM, 64 MM, Silicone, Sterile, Item 431217.
- Z-1657-2020 · TerminatedBIOMET StageOne, Shoulder Cement Spacer Mold 6 MM, 42 X 18 X 46 MM, Silicone, Sterile, Item 431406.
- Z-1658-2020 · TerminatedBIOMET StageOne, Shoulder Cement Spacer Mold 8 MM, 46 X 18 X 53 MM, Silicone, Sterile, Item 431408.
- Z-1660-2020 · TerminatedBIOMET StageOne, Shoulder Cement Spacer Mold 12 MM, 54 X 21 X 64 MM, Silicone, Sterile, Item 431412.
- Z-1661-2020 · TerminatedBIOMET StageOne, Shoulder Cement Spacer Mold 14 MM, 58 X 24 X 64 MM, Silicone, Sterile, Item 431414.
- Z-1662-2020 · TerminatedBIOMET StageOne Knee Femoral Cement Spacer Mold, 60 MM, Silicone, Sterile, Item 432160.
- Z-1663-2020 · TerminatedBIOMET StageOne Knee Femoral Cement Spacer Mold, 65 MM, Silicone, Sterile, Item 432165.
- Z-1664-2020 · TerminatedBIOMET StageOne Knee Femoral Cement Spacer Mold, 70 MM, Silicone, Sterile, Item 432170.
- Z-1665-2020 · TerminatedBIOMET StageOne Knee Femoral Cement Spacer Mold, 75 MM, Silicone, Sterile, Item 432175.
- Z-1666-2020 · TerminatedBIOMET StageOne Knee Tibial Cement Spacer Mold, 65 MM, Silicone, Sterile, Item 433165.
- Z-1667-2020 · TerminatedBIOMET StageOne Knee Tibial Cement Spacer Mold, 70 MM, Silicone, Sterile, Item 433170.
- Z-1668-2020 · TerminatedBIOMET StageOne Knee Tibial Cement Spacer Mold, 75 MM, Silicone, Sterile, Item 433175.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page