ARCHITECT iGentamicin Reagent Kit - Product Usage: The… recallZ-2117-2020 · Abbott Gmbh & Co. KG · initiated Apr 24, 2020

Class II Open, Classified All recalls posted in May 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Gmbh & Co. KG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.

FDA root cause category: Under Investigation by firm

What the company is doing

On April 24, 2020, Abbott issued a Product Correction letter to all customers who received shipments of the three impacted lots of ARCHITECT iGenatamicin Reagent kits. U.S. customers were sent the letter via FedEx priority overnight express. In the U.S., the customers will be requested to return to Abbott a Customer Reply Form acknowledging the receipt or a request for assistance.

Recall details

Recall number
Z-2117-2020
Recall class
Class II
Status
Open, Classified
Initiated by the company
Apr 24, 2020
Posted by the FDA
May 12, 2020
Recalling firm
Abbott Gmbh & Co. KG
Quantity
6087 Reagent Kits
Product code
LCD
FDA submissions
K102699
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CT, DE, FL, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MT, ND, NE, NJ, NV, NY, OH, OK, PA, RI, SC, TN,TX, UT, VA, VT, WA, WI, PR. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Belarus, Belgium, Brazil, Brunei, Canada, Cayman Islands, Chile, China, Costa Rica, Croatia, Curacao, Czech Republic, Faroe Islands, France, Germany, Greece, Greenland, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Lithuania, Malaysia, Mauritius, Namibia, Netherlands, New Zealand, Norway, Oman, Palestinian Territory, Philippines, Poland, Portugal, Republic of Yemen, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tunisia, UAE, United Kingdom and Uruguay.

Product description

ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.

Code information (models, lots, serial numbers)

MODEL: 1P31-25 LOT NUMBERS: 04281BE00, 08510BE00 and 11221BE00 Shelf life per LOT: 04281BE00;2020-MAY-21 08510BE00; 2020-AUG-08 11221BE00;2020-NOV-18

Other devices under the same FDA submission

1 device cleared under K102699 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates