Atellica CH Fructosamine Reagent (FRUC), Siemens Material… recallZ-2281-2020 · Siemens Healthcare Diagnostics, Inc. · initiated May 4, 2020
Class II Terminated All recalls posted in June 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
FDA root cause category: Device Design
What the company is doing
On May 4, 2020, Siemens notified their customers that use the Fructosamine and Enzymatic Creatinine assays on the Atellica, ADVIA, and Dimension platforms of the potential for falsely depressed results in the presence of NAPQI. An Urgent Medical Device Correction Notice was sent to consignees in the US and an Urgent Field Safety Notice was sent to customers outside the US. The letter stated that the Limitations of the Procedure section of the Instructions for Use (IFU) will be updated with specific information related to the assay/platform. Customers were asked to complete and return the Effectiveness Check questionnaire attached to the letter within 30 days. In addition, customers were informed that field service personnel will be sent an Service Knowledge Base (SKB) describing the issue and instructing them how to deal with customer questions.
Recall details
- Recall number
- Z-2281-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 4, 2020
- Posted by the FDA
- Jun 5, 2020
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 26
- Product code
- LCP
- FDA submissions
- K131307
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.
Product description
Atellica CH Fructosamine Reagent (FRUC), Siemens Material Number 11097637, UDI Number 00630414595580 use in the quantitative measurement of glycated protein (fructosamine) in human serum and plasma for monitoring the blood glucose levels of diabetic patients over the preceding two-three weeks.
Code information (models, lots, serial numbers)
Lot # 291011 ***Added 9/9/20*** 201115
Other devices under the same FDA submission
2 devices cleared under K131307 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Atellica CH Fruc 11097637Siemens Healthcare Diagnostics Inc. · ASSAY, GLYCOSYLATED HEMOGLOBIN2 recalls0 reports
- ADVIA® Chemistry FRUC Reagents 10361941Siemens Healthcare Diagnostics Inc. · ASSAY, GLYCOSYLATED HEMOGLOBIN1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85602).
- Z-2280-2020 · TerminatedADVIA Chemistry Fructosamine Reagent (FRUC), Siemens Material Number 10361941, UDI Number 00630414558073 - Product Usage: use in the quantitative measurement of glycated protein (f…
- Z-2282-2020 · TerminatedADVIA Chemistry Enzymatic Creatinine_2, Siemens Material Number 10335869, UDI Number 00630414526690 - Product Usage: use in the quantitative determination of creatinine in human se…
- Z-2283-2020 · TerminatedAtellica CH Enzymatic Creatinine_2, Siemens Material Number 11097533, UDI Number 00630414595795 - Product Usage: use in the quantitative determination of creatinine in human serum,…
- Z-2284-2020 · TerminatedDimension Enzymatic Creatinine, Siemens Material Number 10471520, UDI Number 00842768030512 - Product Usage: use in the quantitative determination of creatinine in human serum, pla…
- Z-2285-2020 · TerminatedDimension Vista Enzymatic Creatinine, Siemens Material Number 10700444, UDI Number 00842768036262 - Product Usage: use in the quantitative determination of creatinine in human seru…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page