Skip to content

Dimension Vista Enzymatic Creatinine, Siemens Material… recallZ-2285-2020 · Siemens Healthcare Diagnostics, Inc. · initiated May 4, 2020

Class II Terminated All recalls posted in June 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.

FDA root cause category: Device Design

What the company is doing

On May 4, 2020, Siemens notified their customers that use the Fructosamine and Enzymatic Creatinine assays on the Atellica, ADVIA, and Dimension platforms of the potential for falsely depressed results in the presence of NAPQI. An Urgent Medical Device Correction Notice was sent to consignees in the US and an Urgent Field Safety Notice was sent to customers outside the US. The letter stated that the Limitations of the Procedure section of the Instructions for Use (IFU) will be updated with specific information related to the assay/platform. Customers were asked to complete and return the Effectiveness Check questionnaire attached to the letter within 30 days. In addition, customers were informed that field service personnel will be sent an Service Knowledge Base (SKB) describing the issue and instructing them how to deal with customer questions.

Recall details

Recall number
Z-2285-2020
Recall class
Class II
Status
Terminated
Initiated by the company
May 4, 2020
Posted by the FDA
Jun 5, 2020
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
23,824
Product code
JFY
FDA submissions
K090330
Distribution
Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.

Product description

Dimension Vista Enzymatic Creatinine, Siemens Material Number 10700444, UDI Number 00842768036262 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.

Code information (models, lots, serial numbers)

Lot #s 20027AC, 19336AA, 19273AA, 19218AB, 19192AB, 19189AA, 19175AA, ***Updated 8/4/20*** 20083AA   ***Updated 11/2/20*** 20128AB 20202AA

Other devices under the same FDA submission

1 device cleared under K090330 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 85602).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page