Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product… recallZ-2408-2020 · Arrow International Inc · initiated May 20, 2020
Class I Terminated All recalls posted in June 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
FDA root cause category: Component design/selection
What the company is doing
Urgent Medical Device Correction notification letters dated 5/20/20 were sent to customers. ACTIONS TO BE TAKEN BY FACILITIES (Immediate): Please immediately check your inventory of Arrow AutoCAT2 and Arrow AC3 Optimus IABPs, whether stored or in use, and determine if you have an IABP with a model number listed above. Please see Appendix 1 for product codes and serial numbers. If an IABP in the scope of this Medical Device Correction displays a System Error 3 or High Baseline alarm, now or at any point in the future until the Long-Term Corrective Action detailed below occurs, immediately quarantine the device and contact Teleflex at 1-855-419-8507 or [email protected] to receive support for inspection and servicing of the impacted device. We recommend against using an IABP for ground or air transport between medical facilities pending implementation of the Customer Action (Long Term). For IABP use during ground or air transport prior to implementation of the Customer Action (Long Term), it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered. Customer Action (Subsequent): Teleflex will contact each impacted facility to schedule inspection and servicing of all IABP units that contain this internal component. This inspection and servicing will be performed onsite at facilities, for all IABP units. Facilities should adhere to the following instructions pending implementation of the Teleflex Long-Term Corrective Action described later in this letter: 1. Ensure that a backup IABP is available as instructed within the Operator Manual. If no such replacement IABP is immediately available, it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered. 2. IABP units should be closely monitored during delivery of IAB
Recall details
- Recall number
- Z-2408-2020
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- May 20, 2020
- Posted by the FDA
- Jun 26, 2020
- Recalling firm
- Arrow International Inc
- Quantity
- 2123 total
- Product code
- DSP
- FDA submissions
- K060309
- Distribution
- Worldwide distribution.
Product description
Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Serial number ranges: Arrow AutoCAT 2 Intra-Aortic Balloon Pump AUTOCAT 2 IAP-0400 140868V 140901V through 140909V 140914V through 140920V 150102V 150106V through 150107V 150110V through 150113V 150115V through 150116V 150220V 150401V 150407V 150409V 150507V through 150513V 150601V through 150605V 150608V through 150611V 150625V 150702V through 150705V 150707V 150720V 150801V through 150802V 150821V 150824V 150827V through 150830V 150901V through 150906V 150908V through 150912V 150919V through 150923V 150925V through 150939V 151009V through 151013V 151015V through 151016V 151019V 151021V through 151027V 151029V 151102V through 151103V 151105V 151108V 151111V 151113V 151201V through 151208V 151210V through 151212V 151214V through 151218V Arrow AutoCAT 2 Intra-Aortic Balloon Pump AUTOCAT 2 IAP-0400 151220V 160101V through 160103V 160105V through 160109V 160112V through 160118V 160120V 160122V 160126V 160201V through 160209V 160211V 160213V through 160214V 160216V through 160217V 160219V 160312V through 160314V 160316V through 160323V 160325V through 160327V 160331V 160401V through 160410V 160412V 160418V through 160422V 160519V 160525V 160529V through 160530V 160605V 160614V 160618V 160630V 160638V through 160639V 160645V 160703V 160718V 160720V 160726V 160728V through 160730V 170101V through 170110V 170401V through 170410V 170501V through 170520V 170601V through 170625V 170811V through 170830V 171021V through 171030V 171037V through 171046V 171106V through 171115V 180131V through 180150V Arrow AutoCAT 2 Intra-Aortic Balloon Pump AUTOCAT 2 IAP-0400 180211V through 180230V 180411V through 180420V 180441V through 180450V 180521V through 180530V 180621V through 180630V 180711V through 180720V 180741V through 180750V 180811V through 180820V 181011V through 181040V 181111V through 181130V 190121V through 190130V 190141V through 190150V 190211V through 190220V 190411V through 190420V 190431V through 190440V 190531V through 190540V 190701V through 190712V 190737V through 190748V 190813V through 190824V 190837V through 190848V 190913V through 190924V 190937V through 190948V 191001V through 191012V 191025V through 191036V 191101V through 191112V 191137V through 191148V 200113V through 200136V 200314V through 200324V
Other devices under the same FDA submission
11 devices cleared under K060309 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ARROW IPN000320Teleflex Incorporated · System, balloon, intra-aortic and control2 deaths383 reports
- ARROW IPN000302Teleflex Incorporated · System, balloon, intra-aortic and control54 reports
- ARROW IPN000325Teleflex Incorporated · System, balloon, intra-aortic and control44 reports
- ARROW IPN000324Teleflex Incorporated · System, balloon, intra-aortic and control14 reports
- ARROW IPN000329Teleflex Incorporated · System, balloon, intra-aortic and control8 reports
- ARROW IPN000308Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000312Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000333Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000334Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000323Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000328Teleflex Incorporated · System, balloon, intra-aortic and control1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 85684).
- Z-2409-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 SPANISH, Product Code IAP-0400E
- Z-2410-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE, Product Code IAP-0500
- Z-2411-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE SPANISH, Product Code IAP-0500E
- Z-2412-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE ITALIAN, Product Code IAP-0500I
- Z-2413-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J
- Z-2414-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE DUTCH, Product Code IAP-0500NL
- Z-2415-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page