Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE… recallZ-2413-2020 · Arrow International Inc · initiated May 20, 2020
Class I Terminated All recalls posted in June 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
FDA root cause category: Component design/selection
What the company is doing
Urgent Medical Device Correction notification letters dated 5/20/20 were sent to customers. ACTIONS TO BE TAKEN BY FACILITIES (Immediate): Please immediately check your inventory of Arrow AutoCAT2 and Arrow AC3 Optimus IABPs, whether stored or in use, and determine if you have an IABP with a model number listed above. Please see Appendix 1 for product codes and serial numbers. If an IABP in the scope of this Medical Device Correction displays a System Error 3 or High Baseline alarm, now or at any point in the future until the Long-Term Corrective Action detailed below occurs, immediately quarantine the device and contact Teleflex at 1-855-419-8507 or [email protected] to receive support for inspection and servicing of the impacted device. We recommend against using an IABP for ground or air transport between medical facilities pending implementation of the Customer Action (Long Term). For IABP use during ground or air transport prior to implementation of the Customer Action (Long Term), it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered. Customer Action (Subsequent): Teleflex will contact each impacted facility to schedule inspection and servicing of all IABP units that contain this internal component. This inspection and servicing will be performed onsite at facilities, for all IABP units. Facilities should adhere to the following instructions pending implementation of the Teleflex Long-Term Corrective Action described later in this letter: 1. Ensure that a backup IABP is available as instructed within the Operator Manual. If no such replacement IABP is immediately available, it is recommended that the risks and benefits of using the IABP be assessed by the medical team treating the patient and that alternative circulatory support devices be considered. 2. IABP units should be closely monitored during delivery of IAB
Recall details
- Recall number
- Z-2413-2020
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- May 20, 2020
- Posted by the FDA
- Jun 26, 2020
- Recalling firm
- Arrow International Inc
- Quantity
- 2123 total
- Product code
- DSP
- FDA submissions
- K060309
- Distribution
- Worldwide distribution.
Product description
Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J
Code information (models, lots, serial numbers)
Serial Number Ranges: AUTOCAT2 WAVE JAPANESE IAP-0500J 140866W through 140867W 150117W through 150118W 150120W 150201W 150305W 150307W through 150309W 150623W through 150624W 150713W through 150716W 151114W 151209W 151213W 151219W 160503W 160511W through 160515W 160517W 160635W through 160636W 160640W through 160641W 160731W through 160733W 171105W 171117W 171119W through 171120W 171201W 180120W 180301W
Other devices under the same FDA submission
11 devices cleared under K060309 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ARROW IPN000320Teleflex Incorporated · System, balloon, intra-aortic and control2 deaths383 reports
- ARROW IPN000302Teleflex Incorporated · System, balloon, intra-aortic and control54 reports
- ARROW IPN000325Teleflex Incorporated · System, balloon, intra-aortic and control44 reports
- ARROW IPN000324Teleflex Incorporated · System, balloon, intra-aortic and control14 reports
- ARROW IPN000329Teleflex Incorporated · System, balloon, intra-aortic and control8 reports
- ARROW IPN000308Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000312Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000333Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000334Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000323Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000328Teleflex Incorporated · System, balloon, intra-aortic and control1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 85684).
- Z-2408-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400
- Z-2409-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 SPANISH, Product Code IAP-0400E
- Z-2410-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE, Product Code IAP-0500
- Z-2411-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE SPANISH, Product Code IAP-0500E
- Z-2412-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE ITALIAN, Product Code IAP-0500I
- Z-2414-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE DUTCH, Product Code IAP-0500NL
- Z-2415-2020 · TerminatedArrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page