Port Access Kit recallZ-2662-2020 · Becton Dickinson & Company · initiated Jun 19, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
FDA root cause category: Process control
What the company is doing
URGENT MEDICAL DEVICE RECALL notification letters dated 6/19/20 were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Immediately review your inventory for the specific catalog and lot numbers referenced in Attachment A. Discontinue use of those devices and destroy all product subject to this recall following your institutions process for destruction. MDS-20-1979 Page 2 of 28 BD Medical 1 Becton Drive Franklin Lakes, NJ 07417 USA bd.com 2. Share this notification with all users of the product within your facility network to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your credit accordingly for the discarded material. a. Customers who purchased product directly from BD will receive credit. b. Customers who did not purchase product directly from BD will need to contact their distributor to obtain credit. 4. Report any adverse health consequences experienced with the use of this product to BD at [email protected]. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1800FDA1088 (18003321088) Mail: MedWatch, HF2, FDA, 5600 Fishers Lane, Rockville, MD 208529787 Customer/Technical Support Phone: 8002901689 Monday Friday between 8:00 am to 5:00 pm MDT [email protected] An Urgent Medical Device Recall notification letter dated 8/25/20 was sent to customers informing them of two additional lots of affected product. An URGENT MEDICAL DEVICE RECALL UPDATE dated 10/7/20 was sent to customers informing them of two additional lots of affected product.
Recall details
- Recall number
- Z-2662-2020
- Status
- Terminated
- Initiated by the company
- Jun 19, 2020
- Posted by the FDA
- Jul 20, 2020
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 5,296,132 total
- Product code
- PTI
- FDA submissions
- K153440
- Distribution
- Worldwide Distribution
Product description
PowerLoc MAX Power Injectable Infusion Port Access Kit; Catalog Numbers 2131910 (UDI 00801741044151), 2131915 (UDI 00801741044168), 2131975 (UDI 00801741044175), 2132010 (UDI 00801741044151), 2132015 (UDI 00801741044199), 2132075 (UDI 00801741044205), 2132210 (UDI 00801741044212), 2132275 (UDI 00801741044236), 2141910 (UDI 00801741044243), 2141915 (UDI 00801741044250), 2141975 (UDI 00801741044267), 2142010 (UDI 00801741044274), 2142015 (UDI 00801741044281), 2142075 (UDI 00801741044298), 2142210 (UDI 00801741044304), 2142215 (UDI 00801741044311), 2142275 (UDI 00801741044328)
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Catalog Numbers 2131910 (UDI 00801741044151), 2131915 (UDI 00801741044168), 2131975 (UDI 00801741044175), 2132010 (UDI 00801741044151), 2132015 (UDI 00801741044199), 2132075 (UDI 00801741044205), 2132210 (UDI 00801741044212), 2132275 (UDI 00801741044236), 2141910 (UDI 00801741044243), 2141915 (UDI 00801741044250), 2141975 (UDI 00801741044267), 2142010 (UDI 00801741044274), 2142015 (UDI 00801741044281), 2142075 (UDI 00801741044298), 2142210 (UDI 00801741044304), 2142215 (UDI 00801741044311), 2142275 (UDI 00801741044328). Lot Numbers: REDW3798 REDY2866 REDZ1719 REEN3137 REEP0951 REEQ2684 REEQ4092 REER1229 REEQ0664 REEQ1649 REER0459 REDZ1818 REEP3049 REDW1092 REDX3754 REDY1682 REDY2971 REDY3530 REDZ1070 REEN3271 REEN4480 REEP0992 REEP3957 REEQ1647 REEQ2724 REEQ4093 REER1266 REDY3719 REEN1270 REEP0952 REEQ0684 REEQ2676 REEQ4054 REER0460 REDT2879 REDU2256 REDU3371 REDU4487 REDV1088 REDV2972 REDV3470 REDW0402 REDW1087 REDX1518 REDY1804 REDY2867 REDY3531 REDZ1069 REEN1104 REEN4481 REEP0953 REEP3958 REEQ0665 REEQ1650 REER0530 REEN2139 REEP4053 REER0461 REDY3720 REEQ4094 REDW3682 REDY0827 REDZ1071 REEP0994 REER0539 REEN4551 REEP0954 REEQ1648 REDW1197 REDY2972 REDZ3130 REEP0995 REEP2909 REDW0546 REDX1519 REDX4930 REDY1683 REDY2868 REDY3532 REDZ1064 REDZ1819 REEN1046 REEN2146 REEP0955 REEP3052 REEP3959 REER0462 REDZ1721 REEQ2678 REDV1236 REDW0407 REDW0550 REDX4852 REDY0773 REDY0828 REDY1684 REDY2973 REDZ1063 REDZ1072 REDZ1820 REEN1271 REEN3273 REEP0956 REEP0996 REEP1728 REEP3053 REEP4054 REEQ1664 REEQ2720 REER0540 REDW3683 REDY3721 REEN4552 REER1261 REEN3155 REDW0403 REDY1685 REDZ1722 REEP0997 REEQ0666
Devices named in this recall
16 devices in the FDA device registry whose identifier appears in the recall's code information.
- Port Access Kit 2142075Bard Access Systems, Inc. · Set, administration, intravascular1 recall28 reports
- Port Access Kit 2132075Bard Access Systems, Inc. · Set, administration, intravascular1 recall3 reports
- Port Access Kit 2141975Bard Access Systems, Inc. · Set, administration, intravascular1 recall1 report
- Port Access Kit 2131910Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2131915Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2131975Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2132015Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2132210Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2132275Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2141910Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2141915Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2142010Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2142015Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2142210Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2142215Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2142275Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 85853).
- Z-2660-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Set AllPoints Port Access System; Catalog Numbers AN132075 (UDI 00801741118968), AN142010 (UDI 00801741119231), AN142010G (UDI 00801741119323…
- Z-2661-2020 · TerminatedPowerLoc MAX Power-Injectable Infusion Set myPICK Port Access Kit; Catalog Numbers CPA00001 (UDI 00801741098345), CPA00002 (UDI 00801741098352), CPA00028 (UDI 00801741129728), CPA0…
- Z-2663-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers 2132010G (UDI 00801741095504), 2132015G (UDI 00801741095…
- Z-2664-2020 · TerminatedSafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI…
- Z-2665-2020 · TerminatedSafeStep Huber Needle Set myPICK Port Access Kit; Catalog Numbers CPA00020A (UDI 00801741129827), CPA00042A (UDI 00801741182969), CPA00049 (UDI 00801741175695)
- Z-2666-2020 · TerminatedSafeStep Huber Needle Set Port Access Kit; Catalog Numbers PA-0029 (UDI 00801741085482), PA-0029YN (UDI 00801741085499), PA-0030 (UDI 00801741085505), PA-0030YN (UDI 00801741085512…
- Z-2667-2020 · TerminatedSafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing; Catalog Number CPA00040 (UDI 00801741143892)
- Z-2668-2020 · TerminatedSafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G …
- Z-2669-2020 · TerminatedSafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI…
- Z-2670-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Set; Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336), 0131975 (UDI 00801741047343), 0132010 (UDI 00801741047350),…
- Z-2671-2020 · TerminatedSafeStep Huber Needle Set; Catalog Numbers LH-0029 (UDI 00801741066160), LH-0030 (UDI 00801741066184), LH-0031 (UDI 00801741066207), LH-0032 (UDI 00801741066221), LH-0033 (UDI 008…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page