Port Access Kit recallZ-2663-2020 · Becton Dickinson & Company · initiated Jun 19, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
FDA root cause category: Process control
What the company is doing
URGENT MEDICAL DEVICE RECALL notification letters dated 6/19/20 were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Immediately review your inventory for the specific catalog and lot numbers referenced in Attachment A. Discontinue use of those devices and destroy all product subject to this recall following your institutions process for destruction. MDS-20-1979 Page 2 of 28 BD Medical 1 Becton Drive Franklin Lakes, NJ 07417 USA bd.com 2. Share this notification with all users of the product within your facility network to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your credit accordingly for the discarded material. a. Customers who purchased product directly from BD will receive credit. b. Customers who did not purchase product directly from BD will need to contact their distributor to obtain credit. 4. Report any adverse health consequences experienced with the use of this product to BD at [email protected]. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1800FDA1088 (18003321088) Mail: MedWatch, HF2, FDA, 5600 Fishers Lane, Rockville, MD 208529787 Customer/Technical Support Phone: 8002901689 Monday Friday between 8:00 am to 5:00 pm MDT [email protected] An Urgent Medical Device Recall notification letter dated 8/25/20 was sent to customers informing them of two additional lots of affected product. An URGENT MEDICAL DEVICE RECALL UPDATE dated 10/7/20 was sent to customers informing them of two additional lots of affected product.
Recall details
- Recall number
- Z-2663-2020
- Status
- Terminated
- Initiated by the company
- Jun 19, 2020
- Posted by the FDA
- Jul 20, 2020
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 5,296,132 total
- Product code
- PTI
- FDA submissions
- K153440
- Distribution
- Worldwide Distribution
Product description
PowerLoc MAX Power Injectable Infusion Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers 2132010G (UDI 00801741095504), 2132015G (UDI 00801741095511), 2141910G (UDI 00801741095542), 2142010G (UDI 00801741085680), 2142015G (UDI 00801741085680), 2142210G (UDI 00801741085680)
Code information (models, lots, serial numbers)
Catalog Numbers 2132010G (UDI 00801741095504), 2132015G (UDI 00801741095511), 2141910G (UDI 00801741095542), 2142010G (UDI 00801741085680), 2142015G (UDI 00801741085680), 2142210G (UDI 00801741085680). Lot Numbers: REDY0771 REDY2980 REEN1111 REEN4550 REER0529 REEN2070 REEP0993 REDX0120 REDW1093 REDW2530 REDX1388 REDY0772 REEN1112 REEN2779 REEP2910 REEQ4095 REER2358 REDX0188
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Port Access Kit 2142010GBard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2132010GBard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2132015GBard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access Kit 2141910GBard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 85853).
- Z-2660-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Set AllPoints Port Access System; Catalog Numbers AN132075 (UDI 00801741118968), AN142010 (UDI 00801741119231), AN142010G (UDI 00801741119323…
- Z-2661-2020 · TerminatedPowerLoc MAX Power-Injectable Infusion Set myPICK Port Access Kit; Catalog Numbers CPA00001 (UDI 00801741098345), CPA00002 (UDI 00801741098352), CPA00028 (UDI 00801741129728), CPA0…
- Z-2662-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Port Access Kit; Catalog Numbers 2131910 (UDI 00801741044151), 2131915 (UDI 00801741044168), 2131975 (UDI 00801741044175), 2132010 (UDI 00801…
- Z-2664-2020 · TerminatedSafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI…
- Z-2665-2020 · TerminatedSafeStep Huber Needle Set myPICK Port Access Kit; Catalog Numbers CPA00020A (UDI 00801741129827), CPA00042A (UDI 00801741182969), CPA00049 (UDI 00801741175695)
- Z-2666-2020 · TerminatedSafeStep Huber Needle Set Port Access Kit; Catalog Numbers PA-0029 (UDI 00801741085482), PA-0029YN (UDI 00801741085499), PA-0030 (UDI 00801741085505), PA-0030YN (UDI 00801741085512…
- Z-2667-2020 · TerminatedSafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing; Catalog Number CPA00040 (UDI 00801741143892)
- Z-2668-2020 · TerminatedSafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G …
- Z-2669-2020 · TerminatedSafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI…
- Z-2670-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Set; Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336), 0131975 (UDI 00801741047343), 0132010 (UDI 00801741047350),…
- Z-2671-2020 · TerminatedSafeStep Huber Needle Set; Catalog Numbers LH-0029 (UDI 00801741066160), LH-0030 (UDI 00801741066184), LH-0031 (UDI 00801741066207), LH-0032 (UDI 00801741066221), LH-0033 (UDI 008…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page