Port Access recallZ-2664-2020 · Becton Dickinson & Company · initiated Jun 19, 2020
Class II Terminated All recalls posted in July 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
FDA root cause category: Process control
What the company is doing
URGENT MEDICAL DEVICE RECALL notification letters dated 6/19/20 were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Immediately review your inventory for the specific catalog and lot numbers referenced in Attachment A. Discontinue use of those devices and destroy all product subject to this recall following your institutions process for destruction. MDS-20-1979 Page 2 of 28 BD Medical 1 Becton Drive Franklin Lakes, NJ 07417 USA bd.com 2. Share this notification with all users of the product within your facility network to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your credit accordingly for the discarded material. a. Customers who purchased product directly from BD will receive credit. b. Customers who did not purchase product directly from BD will need to contact their distributor to obtain credit. 4. Report any adverse health consequences experienced with the use of this product to BD at [email protected]. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1800FDA1088 (18003321088) Mail: MedWatch, HF2, FDA, 5600 Fishers Lane, Rockville, MD 208529787 Customer/Technical Support Phone: 8002901689 Monday Friday between 8:00 am to 5:00 pm MDT [email protected] An Urgent Medical Device Recall notification letter dated 8/25/20 was sent to customers informing them of two additional lots of affected product. An URGENT MEDICAL DEVICE RECALL UPDATE dated 10/7/20 was sent to customers informing them of two additional lots of affected product.
Recall details
- Recall number
- Z-2664-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 19, 2020
- Posted by the FDA
- Jul 20, 2020
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 5,296,132 total
- Product code
- PTI
- FDA submissions
- K153440
- Distribution
- Worldwide Distribution
Product description
SafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI 00801741119446), AN162075 (UDI 00801741119675), AN162075G (UDI 00801741119439)
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI 00801741119446), AN162075 (UDI 00801741119675), AN162075G (UDI 00801741119439). Lot Numbers: REDT0554 REDW1158 REDW2451 REDX1491 REDY1817 REDY2798 REEN3076 REEP3245 REEP3545 REDR2659 REDS4631 REDT0709 REDT4233 REDU2308 REDV1300 REDV3619 REDX2864 REDY3766 REEP3542 REEQ0644 REER1047 REDX4729 REDZ0700 REDZ1905 REEN1900 REEN2797 REEP3240 REEP4079 REEQ0719 REEQ1251 REER1105 REEQ1250 REDR2870 REDS0199 REDS3479 REDT1592 REDU0734 REDU1349 REDU2234 REDU3360 REDU4418 REDV1066 REDV3684 REDW0388 REDW1159 REDY1595 REDY2936 REDY3545 REDZ1827 REEN2784 REEP0777 REER0610 REER2163 REDS4692 REDT0555 REDU1332 REDU2236 REDV1956 REDW3670 REDX1492 REDX3881 REDX4827 REDY1818 REDY2791 REDY3767 REDZ0388 REDZ1906 REDZ3147 REEN3311 REEQ1172 REER1106 REER2125
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- Port Access AN162075Bard Access Systems, Inc. · Set, administration, intravascular1 recall3 reports
- Port Access AN162075GBard Access Systems, Inc. · Set, administration, intravascular1 recall2 reports
- Port Access AN162010Bard Access Systems, Inc. · Set, administration, intravascular1 recall1 report
- Port Access AN162010GBard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access AN152075Bard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
- Port Access AN152075GBard Access Systems, Inc. · Set, administration, intravascular1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 85853).
- Z-2660-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Set AllPoints Port Access System; Catalog Numbers AN132075 (UDI 00801741118968), AN142010 (UDI 00801741119231), AN142010G (UDI 00801741119323…
- Z-2661-2020 · TerminatedPowerLoc MAX Power-Injectable Infusion Set myPICK Port Access Kit; Catalog Numbers CPA00001 (UDI 00801741098345), CPA00002 (UDI 00801741098352), CPA00028 (UDI 00801741129728), CPA0…
- Z-2662-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Port Access Kit; Catalog Numbers 2131910 (UDI 00801741044151), 2131915 (UDI 00801741044168), 2131975 (UDI 00801741044175), 2132010 (UDI 00801…
- Z-2663-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers 2132010G (UDI 00801741095504), 2132015G (UDI 00801741095…
- Z-2665-2020 · TerminatedSafeStep Huber Needle Set myPICK Port Access Kit; Catalog Numbers CPA00020A (UDI 00801741129827), CPA00042A (UDI 00801741182969), CPA00049 (UDI 00801741175695)
- Z-2666-2020 · TerminatedSafeStep Huber Needle Set Port Access Kit; Catalog Numbers PA-0029 (UDI 00801741085482), PA-0029YN (UDI 00801741085499), PA-0030 (UDI 00801741085505), PA-0030YN (UDI 00801741085512…
- Z-2667-2020 · TerminatedSafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing; Catalog Number CPA00040 (UDI 00801741143892)
- Z-2668-2020 · TerminatedSafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G …
- Z-2669-2020 · TerminatedSafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI…
- Z-2670-2020 · TerminatedPowerLoc MAX Power Injectable Infusion Set; Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336), 0131975 (UDI 00801741047343), 0132010 (UDI 00801741047350),…
- Z-2671-2020 · TerminatedSafeStep Huber Needle Set; Catalog Numbers LH-0029 (UDI 00801741066160), LH-0030 (UDI 00801741066184), LH-0031 (UDI 00801741066207), LH-0032 (UDI 00801741066221), LH-0033 (UDI 008…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page