Skip to content

Medfusion Syringe Pump Model 4000 recallZ-2734-2020 · Smiths Medical ASD Inc. · initiated Jun 26, 2020

Class I Open, Classified All recalls posted in August 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Inaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.

FDA root cause category: Software design

What the company is doing

Consignees are being contacted via direct mailing and email. Consignees are asked to review and complete the Recall Response form indicating understanding of notification. Distributors are asked to forward the notification to parties to whom product has been distributed. Smiths Medical will send out 2 additional notifications by email. Phone calls will be made to non-responders.

Recall details

Recall number
Z-2734-2020
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Jun 26, 2020
Posted by the FDA
Aug 6, 2020
Recalling firm
Smiths Medical ASD Inc.
Quantity
35616 units
Product code
FRN
FDA submissions
K040899, K111386
Distribution
Worldwide distribution - United Arab Emirates, United States, Afghanistan, Albania, Australia, Belgium, Bermuda, Canada, China, Germany, Denmark, Great Britain, Hong Kong, Indonesia, Ireland, India, Korea, Kuwait, Lebanon, Mexico, Malaysia, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Saudi Arabia, Trinidad and Tobago

Product description

Medfusion Syringe Pump Model 4000; Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1

Code information (models, lots, serial numbers)

Model Numbers: 4000-0105-249, 4000-0105-50, 4000-0105-51, 4000-0105-78, 4000-0106-00, 4000-0106-01, and 4000-0106-231. Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1

Other devices under the same FDA submission

15 devices cleared under K040899, K111386 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 85912).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page