BD SafetyGlideTM 3ml 25G x 1 recallZ-2889-2020 · Becton Dickinson & Company · initiated Jul 28, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Printed scale markings on the 3mL SafetyGlide syringes, with catalog # 305924, Lot# 8289599, are printed incorrectly. The scale is skewed to different degrees, resulting in missing and/or partially printed scale numbers and scale lines.
FDA root cause category: Equipment maintenance
What the company is doing
BD Sent a customer letter dated 28 July 2020. this letter requests customers to review their inventory for effected lots and Destroy all product subject to the recall. The letter requests customers to return the included response form and notify BD of any adverse events associated to this product.
Recall details
- Recall number
- Z-2889-2020
- Status
- Terminated
- Initiated by the company
- Jul 28, 2020
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 125,200 units
- Product code
- FMI
- FDA submissions
- K951254
- Distribution
- US and Canada
Product description
BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack
Code information (models, lots, serial numbers)
Catalog #: 305924 Lot #: 8289599 UDI #: 30382903059240
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BD Luer-Lok, BD SafetyGlide 305924Becton, Dickinson And Company · SYRINGE, ANTISTICK1 recall28 reports
Other devices under the same FDA submission
19 devices cleared under K951254 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- BD SafetyGlide 305916Becton, Dickinson And Company · Needle, hypodermic, single lumen1,767 reports
- BD Safety-Lok 305901Becton, Dickinson And Company · Needle, hypodermic, single lumen176 reports
- McKesson 192-N251SMckesson Medical-Surgical Inc. · Needle, Hypodermic, Single Lumen170 reports
- McKesson 306616Mckesson Medical-Surgical Inc. · Needle, Hypodermic, Single Lumen113 reports
- BD Safety-Lok 305902Becton, Dickinson And Company · Needle, hypodermic, single lumen107 reports
- BD SafetyGlide 305917Becton, Dickinson And Company · Needle, hypodermic, single lumen70 reports
- BD SafetyGlide 305903Becton, Dickinson And Company · Needle, hypodermic, single lumen52 reports
- BD SafetyGlide 305918Becton, Dickinson And Company · Needle, hypodermic, single lumen44 reports
- BD Safety-Lok 305900Becton, Dickinson And Company · Needle, hypodermic, single lumen31 reports
- BD SafetyGlide 305921Becton, Dickinson And Company · Needle, hypodermic, single lumen15 reports
- SafetyGlide Syringe 303327Becton, Dickinson And Company · Needle, hypodermic, single lumen13 reports
- BD SafetyGlide 23x1-1/2 RB TW 304387Becton, Dickinson And Company · Needle, hypodermic, single lumen12 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page