Skip to content

ADVIA Chemistry Systems-Fructosamine (FRUC)-For in vitro… recallZ-0101-2021 · Siemens Healthcare Diagnostics, Inc. · initiated Sep 22, 2020

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia.

FDA root cause category: Device Design

What the company is doing

Siemens Urgent Medical Device Recall (ACHC20-18.A.US.CHC) distributed to customers on 9/22/20, in the United States and an Urgent Field Safety Notice (ACHC20-18.A.OUS.CHC) was distributed to customers outside the United States who have received ADVIA Chemistry Fructosamine in-date reagent lots. For Atellica, an Urgent Medical Device Recall (ACHC20-18.A.US) was distributed to l customers in the United States and an Urgent Field Safety Notice (ACHC20-1.OUS) was distributed to all customers outside the United States who have received Atellica CH in date reagent lots. These letters instruct the customers to: Perform the following actions on the ADVIA Chemistry System 1. Enter the Real-time correction factor in the Analytical Parameters (Chemistry) window. 2. Once the Real-time correction factor has been configured, update QC ranges as described above in the FRUC QC Definition following your laboratory procedures. 3. Update the reference range following your laboratory procedures. 4. Perform a FRUC calibration and process QC. 5. Perform a system back-up. Complete and return the Field Correction Effectiveness Check attached to this letter within 30 days. " Review this letter with your Medical Director. " If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative

Recall details

Recall number
Z-0101-2021
Status
Open, Classified
Initiated by the company
Sep 22, 2020
Posted by the FDA
Oct 13, 2020
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
US: 23 units; OUS: 204 units
Product code
JIT
FDA submissions
K131307
Distribution
CA, DE, MD, NC, NJ, NY, PA Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Brazil¿¿¿¿ Canada¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ P.R.¿China¿¿¿¿¿ Paraguay¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Vietnam¿¿¿¿¿¿¿¿

Product description

ADVIA Chemistry Systems-Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Chemistry systems Siemens Material Number (SMN): 10361941

Code information (models, lots, serial numbers)

All lots UDI: (01)00630414558073(10)501874(17)20201128 (01)00630414558073(10)514594(17)20210221 (01)00630414558073(10)528971(17)20210630

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K131307 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 86504).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page