Atellica CH Systems Fructosamine (FRUC)-For in vitro… recallZ-0102-2021 · Siemens Healthcare Diagnostics, Inc. · initiated Sep 22, 2020
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia.
FDA root cause category: Device Design
What the company is doing
Siemens Urgent Medical Device Recall (ACHC20-18.A.US.CHC) distributed to customers on 9/22/20, in the United States and an Urgent Field Safety Notice (ACHC20-18.A.OUS.CHC) was distributed to customers outside the United States who have received ADVIA Chemistry Fructosamine in-date reagent lots. For Atellica, an Urgent Medical Device Recall (ACHC20-18.A.US) was distributed to l customers in the United States and an Urgent Field Safety Notice (ACHC20-1.OUS) was distributed to all customers outside the United States who have received Atellica CH in date reagent lots. These letters instruct the customers to: Perform the following actions on the ADVIA Chemistry System 1. Enter the Real-time correction factor in the Analytical Parameters (Chemistry) window. 2. Once the Real-time correction factor has been configured, update QC ranges as described above in the FRUC QC Definition following your laboratory procedures. 3. Update the reference range following your laboratory procedures. 4. Perform a FRUC calibration and process QC. 5. Perform a system back-up. Complete and return the Field Correction Effectiveness Check attached to this letter within 30 days. " Review this letter with your Medical Director. " If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative
Recall details
- Recall number
- Z-0102-2021
- Status
- Open, Classified
- Initiated by the company
- Sep 22, 2020
- Posted by the FDA
- Oct 13, 2020
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- US: 28; OUS: 255 units
- Product code
- JIT
- FDA submissions
- K131307
- Distribution
- CA, DE, MD, NC, NJ, NY, PA Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Brazil¿¿¿¿ Canada¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ P.R.¿China¿¿¿¿¿ Paraguay¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Vietnam¿¿¿¿¿¿¿¿
Product description
Atellica CH Systems Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Chemistry systems Siemens Material Number (SMN): 11097637
Code information (models, lots, serial numbers)
All lots UDI: (01)00630414595580(10)201115(17)20210201 (01)00630414595580(10)201207(17)20210601
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH Fruc 11097637Siemens Healthcare Diagnostics Inc. · ASSAY, GLYCOSYLATED HEMOGLOBIN2 recalls0 reports
Other devices under the same FDA submission
1 device cleared under K131307 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ADVIA® Chemistry FRUC Reagents 10361941Siemens Healthcare Diagnostics Inc. · ASSAY, GLYCOSYLATED HEMOGLOBIN1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 86504).
- Z-0101-2021 · Open, ClassifiedADVIA Chemistry Systems-Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Che…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page