Skip to content

YUKON OCT Spinal System recallZ-0520-2021 · K2M, Inc · initiated Nov 3, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from K2M, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Shafts do not pass through the inner cannula of the Anti-Torque Alignment Tube from Lot KTGG due to a manufacturing nonconformance. This could result in a delay of surgery or result in surgical intervention later on.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Our records indicate Stryker previously distributed catalog no. 7601-9001, Lot KTGG to your location in a Stryker Loaner Set. The nonconforming instrument was subsequently identified and quarantined during routine processing of loaner sets by Stryker at its Leesburg facility. This communication is being made for notification purposes only. The only action requested of you is to complete the Business Reply form found on the following page and return it to the email addresses referenced in the form. If you have any questions or concerns, please contact Stryker Regulatory Compliance at 201-749-8090. On behalf of Stryker we thank you sincerely for your help and support in completing the Business Reply Form and returning it to Stryker by November 6th, 2020.

Recall details

Recall number
Z-0520-2021
Status
Terminated
Initiated by the company
Nov 3, 2020
Recalling firm
K2M, Inc
Quantity
4 devices
Product code
NKG
FDA submissions
K193129
Distribution
US Nationwide distribution.

Product description

Yukon OCT Spinal System Yukon Anti-Torque Alignment Tube - Product Usage: The Anti-Torque Alignment Tube (the Alignment Tube ) provides counter rotation to reduce the torque applied to the anatomy when the size 15 Torque limiting shaft (the shaft is used to tighten the Yukon Set Screw to the specified torque.

Code information (models, lots, serial numbers)

Catalog Number/REF 7601-90001 LOT KTGG GTIN 10888857343948

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page