YUKON OCT Spinal System 7601-90001Recalls & reported problems
VB Spine LLC · Anti Torque Alignment Tube · Class 1
All models of YUKON OCT Spinal System (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10888857343948), received between Nov 12, 2020 and Nov 12, 2020 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Nonstandard Device” (7 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 3, 2020Class IITerminated
Shafts do not pass through the inner cannula of the Anti-Torque Alignment Tube from Lot KTGG due to a manufacturing nonconformance. This could result in a delay of surgery or result in surgical intervention later on.
Z-0520-2021 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Nonstandard Device7
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 12, 2020Nonstandard Device; Nonstandard Device
DURING DEVICE INSPECTION, IT WAS NOTED THAT A YUKON ANTI-TORQUE ALIGNMENT TUBE WAS OUT OF SPECIFICATION. WHEN TESTED, THE FINAL TIGHTENING SHAFT WOULD NOT PROPERLY MATE WITH THE ANTI-TORQUE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10833139 ↗MalfunctionNov 12, 2020Nonstandard Device; Nonstandard Device
DURING DEVICE INSPECTION, IT WAS NOTED THAT A YUKON ANTI-TORQUE ALIGNMENT TUBE WAS OUT OF SPECIFICATION. WHEN TESTED, THE FINAL TIGHTENING SHAFT WOULD NOT PROPERLY MATE WITH THE ANTI-TORQUE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10833081 ↗MalfunctionNov 12, 2020Nonstandard Device; Nonstandard Device
DURING DEVICE INSPECTION, IT WAS NOTED THAT A YUKON ANTI-TORQUE ALIGNMENT TUBE WAS OUT OF SPECIFICATION. WHEN TESTED, THE FINAL TIGHTENING SHAFT WOULD NOT PROPERLY MATE WITH THE ANTI-TORQUE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10833075 ↗MalfunctionNov 12, 2020Nonstandard Device; Nonstandard Device
DURING DEVICE INSPECTION, IT WAS NOTED THAT A YUKON ANTI-TORQUE ALIGNMENT TUBE WAS OUT OF SPECIFICATION. WHEN TESTED, THE FINAL TIGHTENING SHAFT WOULD NOT PROPERLY MATE WITH THE ANTI-TORQUE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10833065 ↗MalfunctionNov 12, 2020Nonstandard Device; Nonstandard Device
DURING DEVICE INSPECTION, IT WAS NOTED THAT A YUKON ANTI-TORQUE ALIGNMENT TUBE WAS OUT OF SPECIFICATION. WHEN TESTED, THE FINAL TIGHTENING SHAFT WOULD NOT PROPERLY MATE WITH THE ANTI-TORQUE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10832996 ↗MalfunctionNov 12, 2020Nonstandard Device; Nonstandard Device
DURING DEVICE INSPECTION, IT WAS NOTED THAT A YUKON ANTI-TORQUE ALIGNMENT TUBE WAS OUT OF SPECIFICATION. WHEN TESTED, THE FINAL TIGHTENING SHAFT WOULD NOT PROPERLY MATE WITH THE ANTI-TORQUE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10832913 ↗MalfunctionNov 12, 2020Nonstandard Device; Nonstandard Device
DURING DEVICE INSPECTION, IT WAS NOTED THAT A YUKON ANTI-TORQUE ALIGNMENT TUBE WAS OUT OF SPECIFICATION. WHEN TESTED, THE FINAL TIGHTENING SHAFT WOULD NOT PROPERLY MATE WITH THE ANTI-TORQUE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10832862 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page