VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and… recallZ-1308-2021 · Biomerieux Inc · initiated Jan 28, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomerieux Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).

FDA root cause category: Software design

What the company is doing

The customer letter instructs the customer on how to verify the external communication configuration setting to see if they are impacted by this correction. If impacted, the customer will need to contact their local bioMerieux representative for instructions on how to verify expertised results until a correction is implemented. For a correction, customers can either convert to MYLA or BCI (where this issue will not occur) or continue to verify the expertised results until a software patch is deployed for 9.0x customers.

Recall details

Recall number
Z-1308-2021
Status
Terminated
Initiated by the company
Jan 28, 2021
Recalling firm
Biomerieux Inc
Quantity
6777 systems
Product code
LON
FDA submissions
K050002
Distribution
Worldwide distribution - No US consignees affected and the countries of Andorra, Austria, Bangladesh, Bermuda, Brazil, Brunei Darussalam, El Salvador, French Guiana, French Polynesia, Gibraltar, Greece, Guadeloupe, Guam, Holy See (Vatican City State), Hong Kong, Hungary, India, Indonesia, Ireland, Lao People's Democratic Republic, Luxembourg, Macao, Malaysia, Martinique, Mexico, Myanmar, New Caledonia, New Zealand, Nigeria, Philippines, Reunion, San Marino, Slovakia, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Uruguay.

Product description

VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

VITEK 2 and VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7. Each instrument has unique serial number. (System: VITEK 2 Compact 60, Reference Number: 27560, Catalog Number: C57, UDI: 03573026136796); (System: VITEK 2 Compact 30, Reference Number: 27560R, Catalog Number: C57, UDI: 03573026147501); (System: VITEK 2 Compact 30, Reference Number: 27530, Catalog Number: C57, UDI: 03573026136789); (System: VITEK 2 Compact 30, Reference Number: 27530R, Catalog Number: C57, UDI: 03573026147495); (System: VITEK 2 Compact 15, Reference Number: 27415, Catalog Number: C57, UDI: 03573026220624); (System: VITEK 2 Compact 15, Reference Number: 27415R, Catalog Number: C57, UDI: 03573026220631)

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 87249).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page