VITEK® 2 COMPACT 27530Recalls & reported problems

Biomerieux , Inc. · VITEK® 2 COMPACT 30 is an automated system that performs bacterial ID and AST. A total of 30 ID and/or AST cards can be processed simultaneously per run. · Class 2

All models of VITEK 2 COMPACT (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03573026136789), received between Jan 3, 2020 and Aug 14, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Data Problem” (1 report).

1 FDA recall record name this device's identifier.

5 other recalls cover devices cleared under the same FDA submission (K050002); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 28, 2021Class IITerminated

    bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).

    Z-1308-2021 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Data Problem1
  • Decreased Sensitivity1
  • High Test Results1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • No Device Output1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2026Data Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50257523 ↗
  • MalfunctionDec 1, 2022High Test Results; High Test Results

    INTENDED USE: THE VITEK® 2 SYSTEM IS INTENDED FOR THE AUTOMATED QUANTITATIVE OR QUALITATIVE ANTIMICROBIAL SUSCEPTIBILITY TESTING OF ISOLATED COLONIES FOR MOST CLINICALLY SIGNIFICANT AEROBIC GRAM-NEGATIVE BACILLI, STAPHYLOCOCCUS SPP., ENTEROCOCCUS SPP., STREPTOCOCCUS SPP., AND YEAST. THE VITEK® 2 SYSTEM IS ALSO INTENDED FOR THE AUTOMATED IDENTIFICATION OF MOST CLINICALLY SIGNIFICANT ANAEROBIC ORGANISMS AND CORYNEBACTERIUM SPECIES, FERMENTING AND NON-FERMENTING GRAM-NEGATIVE BACILLI, GRAM-POSITIVE ORGANISMS, FASTIDIOUS ORGANISMS, AND YEASTS AND YEAST-LIKE ORGANISMS. DESCRIPTION OF THE ISSUE: A CUSTOMER IN SPAIN NOTIFIED BIOMERIEUX OF OBTAINING FALSE RESISTANT RESULTS WHEN USING THE VITEK 2 COMPACT 30 - REFERENCE (B)(4) ( SERIAL NUMBER (B)(4)). AT THE MOMENT, IT HAS NOT BEEN NOTIFIED WHICH STRAINS OR ANTIMICROBIALS ARE CONCERNED BY THE FALSE RESISTANT RESULTS. THE CUSTOMER INDICATES THAT ALTERNATE METHOD AND REPEAT TESTING HAVE BEEN PERFORMED TO CONFIRM THE EXPECTED SUSCEPTIBLE RESULTS. THERE IS NO INDICATION OR REPORT FROM THE CUSTOMER THAT THIS EVENT LED TO OR CONTRIBUTED TO ANY DEATH, SERIOUS INJURY, OR SERIOUS DETERIORATION IN THE STATE OF HEALTH FOR THE CONCERNED PATIENT. THIS DI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15892743 ↗
  • MalfunctionOct 26, 2022Decreased Sensitivity; Decreased Sensitivity

    INTENDED USE: THE VITEK® 2 SYSTEM IS INTENDED FOR THE AUTOMATED QUANTITATIVE OR QUALITATIVE ANTIMICROBIAL SUSCEPTIBILITY TESTING OF ISOLATED COLONIES FOR MOST CLINICALLY SIGNIFICANT AEROBIC GRAM-NEGATIVE BACILLI, STAPHYLOCOCCUS SPP., ENTEROCOCCUS SPP., STREPTOCOCCUS SPP., AND YEAST. THE VITEK® 2 SYSTEM IS ALSO INTENDED FOR THE AUTOMATED IDENTIFICATION OF MOST CLINICALLY SIGNIFICANT ANAEROBIC ORGANISMS AND CORYNEBACTERIUM SPECIES, FERMENTING AND NON-FERMENTING GRAM-NEGATIVE BACILLI, GRAM-POSITIVE ORGANISMS, FASTIDIOUS ORGANISMS, AND YEASTS AND YEAST-LIKE ORGANISMS. DESCRIPTION OF THE ISSUE: A CUSTOMER IN INDIA NOTIFIED BIOMÉRIEUX OF HAVING CARD TERMINATION ISSUES WITH THEIR VITEK 2 COMPACT (REF. 27530, SERIAL NUMBER (B)(4) RESULTING IN DELAY IN REPORTING RESULTS FOR TWENTY (20) PATIENT ISOLATES. THE CUSTOMER STATED ALL CARDS LOADED (B)(6) 2022 WERE TERMINATED DUE TO THE OPTICS BEING OUT OF CALIBRATION. INDIA LOCAL CUSTOMER SERVICE (LCS) HAD THE CUSTOMER CLEAN THE OPTICS AND ALSO CONNECTED TO THE SYSTEM VIA REMOTE DIAGNOSTICS TO CHECK THE OPTICS STATUS. THE OPTICS PASSED THE CAL COMMAND; HOWEVER, THE DAC VALUES WERE LOW. THE VALUES DISPLAYED WITH THE READ COMMAND WERE ALSO LOW. LCS D

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15672240 ↗
  • MalfunctionMar 9, 2021No Device Output; No Device Output

    A CUSTOMER IN (B)(6) NOTIFIED BIOMERIEUX OF A MISSING ISOLATE IN ASSOCIATION WITH THE VITEK® 2 COMPACT 30 (REF. 27530, SERIAL NUMBER (B)(4) ), THE VITEK 2 GP ID CARD (REF. 21342, LOT UNKNOWN) AND SOFTWARE VERSION 9.02. THE CUSTOMER OBTAINED A LOW-DISCRIMINATION RESULT AND STATED WHEN THEY TRIED TO OPEN THE ISOLATE RECORD THEY OBTAINED AN ERROR MESSAGE STATED ¿CANNOT FIND THE ISOLATE.¿ THE CUSTOMER RESTARTED THE SYSTEM; THE SAME ERROR MESSAGE WAS OBTAINED WHEN ATTEMPTING TO OPEN THE ISOLATE RECORD. IT IS UNKNOWN IF THE INABILITY TO REVIEW THE ISOLATE RESULTED IN A DELAY OF REPORTING RESULTS. BIOMERIEUX HAS REQUESTED THE CUSTOMER DATA LOGS. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. BIOMERIEUX WILL INITIATE AN INTERNAL INVESTIGATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11447029 ↗
  • MalfunctionJan 3, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEX OF FALSE RESISTANT RESULTS AND DELAYS IN REPORTING RESULTS FOR PATIENT ISOLATES IN ASSOCIATION WITH THE VITEK® 2 COMPACT 30 (REF 27530, SERIAL (B)(4)). THE CUSTOMER STATED THAT THE MAJORITY OF THEIR PATIENT ISOLATES WERE OBTAINING RESISTANT RESULTS ON THE VITEK® 2. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE FALSE RESISTANT RESULTS OR DELAYS LED TO ANY ADVERSE EVENTS RELATED TO ANY PATIENT'S STATE OF HEALTH. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9547285 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K050002. One submission often covers many models, so check the product and lot details in each record.

  • Z-1307-2021 · Jan 28, 2021 · Terminated
    VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short inc
  • Z-0639-2019 · Nov 13, 2018 · Terminated
    VITEK 2 Systems Software Version 9.01 Update Kit.
  • Z-3075-2017 · Jul 12, 2017 · Terminated
    VITEK 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and ye
  • Z-3076-2017 · Jul 12, 2017 · Terminated
    VITEK 2 Compact 30, REF numbers: 27530 and 27530R (clinical), and 27630 and 17630R (industry) software and reagent cards designed for the identification (ID) and antimicrobial sus
  • Z-0991-2013 · Feb 18, 2013 · Terminated
    VITEK 2 Compact computer system when equipped with the following Hewlett-Packard PC models: (1) dc7100 8-port, Cat. #W0441; (2) dc7100 8-port (refurbished), Cat. #W0441R; (3) dc

Other System, Test, Automated, Antimicrobial Susceptibility, Short Incubation devices

All System, Test, Automated, Antimicrobial Susceptibility, Short Incubation devices →