Incisive CT recallZ-1265-2021 · Philips North America Llc · initiated Feb 22, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees
FDA root cause category: Process control
What the company is doing
Philips notified US consignee on 2/22/2021via email including the Field Safety Notice informing of the reason for recall, health risk and action to take: stop using the system until a Philips Field Service engineer arrives on site to implement the correction to lock the gantry tilt function (FCO72800745). After the FCO72800745 implementation, the device can continue to operate without tilt function until the final correction (FCO72800742) is implemented to replace the nonconforming actuators which will remedy the issue and return device to normal use. A Philips Service Engineer will contact you to schedule the Field Actions. Both of the Field Actions FCO72800745 & FCO72800742 will be free of charge.
Recall details
- Recall number
- Z-1265-2021
- Status
- Terminated
- Initiated by the company
- Feb 22, 2021
- Recalling firm
- Philips North America Llc
- Quantity
- 105 units (1 US and 104 OUS)
- Product code
- JAK
- FDA submissions
- K180015
- Distribution
- US Virgin Island Foreign: Australia Bolivia China Germany Hungary Indonesia Italy Japan Korea Korea, Republic of Latvia Palestine Philippines Romania Russian Federation Senegal Thailand Uzbekistan
Product description
Incisive CT Computed Tomography X-ray System Model: 728143
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Serial Numbers: US: 500230 OUS: 500246 500260 500277 500261 530004 530102 530101 530103 530107 530109 530108 530113 530110 530119 530118 530117 530120 530125 530121 530124 530128 530126 530127 530129 530130 530132 530133 530134 530135 530136 530137 530139 530141 530138 530140 530144 530143 530145 530148 530146 530142 530147 530149 530154 530151 530158 530150 530153 530152 530157 530156 530159 530165 530161 530160 530162 530164 530163 530166 530171 530167 530168 530170 530169 530174 530173 530176 530185 530179 530172 530177 530196 530200 530201 530211 530212 530219 530221 500048 500231 500244 500247 500254 500257 500273 500270 500269 500226 500227 500241 500276 500229 500245 500248 500274 500263 500258 500278 500286 500225 500233 500256 500264 500265 UDI: (01)00884838059849(21)304034 (01)00884838059849(21)325078 (01)00884838085015(21)500218 (01)00884838085015(21)500219 (01)00884838085015(21)500221 (01)00884838085015(21)500222 (01)00884838085015(21)500223 (01)00884838085015(21)500224 (01)00884838085015(21)500225 (01)00884838085015(21)500226 (01)00884838085015(21)500227 (01)00884838085015(21)500229 (01)00884838085015(21)500231 (01)00884838085015(21)500233 (01)00884838085015(21)500234 (01)00884838085015(21)500237 (01)00884838085015(21)500238 (01)00884838085015(21)500239 (01)00884838085015(21)500241 (01)00884838085015(21)500242 (01)00884838085015(21)500245 (01)00884838085015(21)500247 (01)00884838085015(21)500248 (01)00884838085015(21)500249 (01)00884838085015(21)500251 (01)00884838085015(21)500252 (01)00884838085015(21)500254 (01)00884838085015(21)500255 (01)00884838085015(21)500256 (01)00884838085015(21)500257 (01)00884838085015(21)500258 (01)00884838085015(21)500261 (01)00884838085015(21)500264 (01)00884838085015(21)500274 (01)00884838085015(21)500283 (01)00884838085015(21)500296 (01)00884838085015(21)530038 (01)00884838085015(21)530107 (01)00884838085015(21)530108 (01)00884838085015(21)530109 (01)00884838085015(21)530113 (01)00884838085015(21)530117 (01)00884838085015(21)530118 (01)00884838085015(21)530119 (01)00884838085015(21)530120 (01)00884838085015(21)530121 (01)00884838085015(21)530124 (01)00884838085015(21)530125 (01)00884838085015(21)530126 (01)00884838085015(21)530127 (01)00884838085015(21)530128 (01)00884838085015(21)530129 (01)00884838085015(21)530130 (01)00884838085015(21)530132 (01)00884838085015(21)530133 (01)00884838085015(21)530134 (01)00884838085015(21)530135 (01)00884838085015(21)530136 (01)00884838085015(21)530137 (01)00884838085015(21)530138 (01)00884838085015(21)530139 (01)00884838085015(21)530140 (01)00884838085015(21)530141 (01)00884838085015(21)530142 (01)00884838085015(21)530143 (01)00884838085015(21)530144 (01)00884838085015(21)530145 (01)00884838085015(21)530146 (01)00884838085015(21)530147 (01)00884838085015(21)530148 (01)00884838085015(21)530149 (01)00884838085015(21)530150 (01)00884838085015(21)530151 (01)00884838085015(21)530152 (01)00884838085015(21)530153 (01)00884838085015(21)530154 (01)00884838085015(21)530156 (01)00884838085015(21)530157 (01)00884838085015(21)530158 (01)00884838085015(21)530159 (01)00884838085015(21)530160 (01)00884838085015(21)530161 (01)00884838085015(21)530162 (01)00884838085015(21)530163 (01)00884838085015(21)530164 (01)00884838085015(21)530165 (01)00884838085015(21)530166 (01)00884838085015(21)530167 (01)00884838085015(21)530168 (01)00884838085015(21)530169 (01)00884838085015(21)530170 (01)00884838085015(21)530171 (01)00884838085015(21)530172 (01)00884838085015(21)530173 (01)00884838085015(21)530174 (01)00884838085015(21)530176 (01)00884838085015(21)530177 (01)00884838085015(21)530179 (01)00884838085015(21)530180 (01)00884838085015(21)530185 (01)00884838085015(21)530196 (01)00884838085015(21)530200 (01)00884838085015(21)530211 (01)00884838085015(21)530219 (01)00884838085015(21)530221
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page