Incisive CTRecalls & reported problems across all models

Philips Healthcare (Suzhou) Co., Ltd. · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

13 open recalls name at least one Incisive CT model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
25
Deaths reported
0
Injuries reported
3
Recalls
16

We found 2 models of Incisive CT by Philips Healthcare (Suzhou) Co., Ltd. in the FDA's device registry (GUDID). Together they are named in 25 adverse event reports in the FDA's MAUDE database, received between Feb 11, 2021 and Mar 26, 2026 — classified as 3 injuries, 22 malfunctions.

The most frequently reported specific device problem is “Intermittent Program or Algorithm Execution” (7 reports).

16 FDA recall records name at least one of these models by its device identifier, 13 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Incisive CT recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 26, 2026Open, Classified

    Potential of failing to trigger clinical scan during Bolus Tracking.

    Z-2988-2026 · Root cause: Radiation Control for Health and Safety Act · names 2 of these models · Recall details · FDA record ↗

  2. Oct 29, 2025Open, Classified

    If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.

    Z-0881-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  3. Sep 25, 2025Open, Classified

    Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.

    Z-0375-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  4. Apr 15, 2024Open, Classified

    Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity

    Z-1872-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. Apr 15, 2024Open, Classified

    Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity

    Z-1871-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. Dec 7, 2023Open, Classified

    If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG signal may lead to the decision to rescan the patient or delay in diagnosis during a clinical emergency.

    Z-0719-2024 · Root cause: Component design/selection · names 2 of these models · Recall details · FDA record ↗

  7. Apr 25, 2023Open, Classified

    Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down

    Z-2258-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Apr 12, 2023Open, Classified

    Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required

    Z-1623-2023 · Root cause: Software design · Recall details · FDA record ↗

  9. Apr 12, 2023Open, Classified

    Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required

    Z-1622-2023 · Root cause: Software design · Recall details · FDA record ↗

  10. Mar 20, 2023Open, Classified

    Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury

    Z-1586-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  11. Mar 20, 2023Open, Classified

    Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury

    Z-1585-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  12. Dec 23, 2022Open, Classified

    Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan. These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operators voice are too quiet, patient cant hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP.

    Z-0640-2024 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  13. Mar 29, 2022Completed

    Three software issues affecting incorrect image display, error interpreting patient images due to image reporting, and patient exposure to incorrect image/function during clinical use.

    Z-0576-2024 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  14. Feb 24, 2021Terminated

    Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position.

    Z-1273-2021 · Root cause: Process control · Recall details · FDA record ↗

  15. Feb 22, 2021Terminated

    Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees

    Z-1265-2021 · Root cause: Process control · Recall details · FDA record ↗

  16. Dec 24, 2020Open, Classified

    Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured

    Z-1507-2024 · Root cause: Software design · Recall details · FDA record ↗

Most reported Incisive CT problems

Device problems

  • Intermittent Program or Algorithm Execution7
  • Program or Algorithm Execution Failure6
  • No Apparent Adverse Event4
  • Use of Device Problem4
  • Detachment of Device or Device Component3
  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions23
  • Laceration(s)2
  • Insufficient Information1
  • No Patient Involvement1

Incisive CT models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Incisive CT adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Philips Healthcare (Suzhou) Co., Ltd. products

All Philips Healthcare (Suzhou) Co., Ltd. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page