Incisive CT recallZ-1273-2021 · Philips Healthcare (Suzhou) Co., Ltd. · initiated Feb 24, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Healthcare (Suzhou) Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position.

FDA root cause category: Process control

What the company is doing

On 2/18/2021, Philips issued an Urgent Medical Device Correction because the firm has detected an issue with some actuators used in the couch of Incisive CT and Access CT systems, which could result in the couch unexpectedly moving slowly downward to its limit position.

Recall details

Recall number
Z-1273-2021
Status
Terminated
Initiated by the company
Feb 24, 2021
Recalling firm
Philips Healthcare (Suzhou) Co., Ltd.
Quantity
109 systems
Product code
JAK
FDA submissions
K180015
Distribution
Worldwide distribution - US Nationwide distribution in the states of KY, FL, St. Croix, and the countries of Australia, Bolivia, China, Germany, Hungary, Italy, Japan, Republic of Korea, Latvia, Palestine, Puerto Rico, Romania, Russia, Saudi Arabia, Senegal, Serbia, Spain, Sweden, Thailand, Uzbekistan.

Product description

Philips CT systems are advanced continuous-rotation computed tomography systems suitable for a wide range of computed tomographic (CT) applications - Product Usage: intended for use in the head and whole body.

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

Unique Device Identifier (01)00884838059849(21)304034  (01)00884838059849(21)325078  (01)00884838085015(21)500218  (01)00884838085015(21)500219  (01)00884838085015(21)500221  (01)00884838085015(21)500222  (01)00884838085015(21)500223  (01)00884838085015(21)500224  (01)00884838085015(21)500225  (01)00884838085015(21)500226  (01)00884838085015(21)500227  (01)00884838085015(21)500229  (01)00884838085015(21)500231  (01)00884838085015(21)500233  (01)00884838085015(21)500234  (01)00884838085015(21)500237  (01)00884838085015(21)500238  (01)00884838085015(21)500239  (01)00884838085015(21)500241  (01)00884838085015(21)500242  (01)00884838085015(21)500245  (01)00884838085015(21)500247  (01)00884838085015(21)500248  (01)00884838085015(21)500249  (01)00884838085015(21)500251  (01)00884838085015(21)500252  (01)00884838085015(21)500254  (01)00884838085015(21)500255  (01)00884838085015(21)500256  (01)00884838085015(21)500257  (01)00884838085015(21)500258  (01)00884838085015(21)500261  (01)00884838085015(21)500264  (01)00884838085015(21)500274  (01)00884838085015(21)500283  (01)00884838085015(21)500296  (01)00884838085015(21)530038  (01)00884838085015(21)530067  (01)00884838085015(21)530071  (01)00884838085015(21)530098  (01)00884838085015(21)530099  (01)00884838085015(21)530101  (01)00884838085015(21)530103  (01)00884838085015(21)530104  (01)00884838085015(21)530106  (01)00884838085015(21)530107  (01)00884838085015(21)530108  (01)00884838085015(21)530109  (01)00884838085015(21)530111  (01)00884838085015(21)530112  (01)00884838085015(21)530113  (01)00884838085015(21)530114  (01)00884838085015(21)530115  (01)00884838085015(21)530116  (01)00884838085015(21)530117  (01)00884838085015(21)530118  (01)00884838085015(21)530119  (01)00884838085015(21)530120  (01)00884838085015(21)530121  (01)00884838085015(21)530124  (01)00884838085015(21)530126  (01)00884838085015(21)530127  (01)00884838085015(21)530128  (01)00884838085015(21)530129  (01)00884838085015(21)530130  (01)00884838085015(21)530132  (01)00884838085015(21)530133 (01)00884838085015(21)530134  (01)00884838085015(21)530135  (01)00884838085015(21)530136  (01)00884838085015(21)530137  (01)00884838085015(21)530138  (01)00884838085015(21)530141  (01)00884838085015(21)530148  (01)00884838085015(21)530150  (01)00884838085015(21)530152  (01)00884838085015(21)530156  (01)00884838085015(21)530157  (01)00884838085015(21)530159  (01)00884838085015(21)530160  (01)00884838085015(21)530163  (01)00884838085015(21)530165  (01)00884838085015(21)530166  (01)00884838085015(21)530167  (01)00884838085015(21)530169  (01)00884838085015(21)530170  (01)00884838085015(21)530171  (01)00884838085015(21)530172  (01)00884838085015(21)530173  (01)00884838085015(21)530174  (01)00884838085015(21)530175  (01)00884838085015(21)530176  (01)00884838085015(21)530177  (01)00884838085015(21)530178  (01)00884838085015(21)530180  (01)00884838085015(21)530181  (01)00884838085015(21)530182  (01)00884838085015(21)530183  (01)00884838085015(21)530186  (01)00884838085015(21)530187  (01)00884838085015(21)530188  (01)00884838085015(21)530189  (01)00884838085015(21)530190  (01)00884838085015(21)530191  (01)00884838085015(21)530192  (01)00884838085015(21)530194  (01)00884838085015(21)530197  (01)00884838085015(21)530198  (01)00884838085015(21)500230

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page