Arrow OnControl Bone Lesion Biopsy System Tray-11 ga 10 cm… recallZ-1374-2021 · Arrow International Inc · initiated Mar 8, 2021
Class II Open, Classified All recalls posted in April 2021
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Difficulty inserting the biopsy ejector rod into the bone lesion biopsy cannula, preventing the biopsy sample from being collected. Use of a defective device is likely to require immediate repeat biopsy, leading to a prolonged procedure
FDA root cause category: Process design
What the company is doing
Teleflex issued to US consignee notification letter on March 8, 2021 via FedEx 2-day mail. Letter states reason for recall, health risk, and action to take: Action list number 1 Medical facilities 1. We request that you check your inventory for product within the scope of this FSCA. Users should cease use and distribution of impacted product and quarantine immediately. 2. If you do not have stock in scope of this FSCA mark the according checkbox on the Acknowledgement Form (Appendix 1) and return the form to the fax number or e-Mail address mentioned below. 3. If you do have stock in scope of this FSCA, mark the according checkbox on the Acknowledgement Form (Appendix 1) and contact customer service by calling the phone number mentioned below. Customer service will issue you with a return number. Write the return number into the respective field in the Acknowledgement Form and return this form immediately to Customer Service. 4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product. Action list number 2 Distributors 1. Provide this field safety notice to all customers who have received product in scope of this FSCA. Your customer is then required to complete the acknowledgement form and return to you. 2. We request that you check your inventory for product within the scope of this FSCA. Cease use and distribution of impacted product and quarantine immediately. You may then return all product in scope. 3. As a distributor, you are then required to confirm to Teleflex that you have completed the field activity outlined above. Upon completion of your actions, please forward the completed Acknowledgement Form to Customer Service. 4. If you have further distributed product outside of your country, please notify Teleflex by return email to the e-Mail address below. Should you require any further information or support concerning this issue, please contact: Customer Service: Contact: Customer Serv
Recall details
- Recall number
- Z-1374-2021
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Mar 8, 2021
- Posted by the FDA
- Apr 7, 2021
- Recalling firm
- Arrow International Inc
- Quantity
- 3,756 units (US); 249 units (OUS) Expanded 810 units: 774 US and 36 OUS
- Product code
- KNW
- FDA submissions
- K113872
- Distribution
- US Nationwide distribution.
Product description
Arrow OnControl Bone Lesion Biopsy System Tray-11 ga 10 cm Access/13 ga 14 cm Biopsy, Product Code: 9464-VC-006 (US) Product Code: 9464-EU-001 (sold in European Regions only) - Product Usage: intended for bone biopsy of the vertebral body and bone lesions.
Code information (models, lots, serial numbers)
Lot Numbers US: 39647326 39647327 39647328 40422734 40422735 Lot Number: 40808740 (EU) EXPANDED RECALL: 5/4/21 Lot Numbers: 38347967 40422740
Other devices under the same FDA submission
3 devices cleared under K113872 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ARROW IPN033806Teleflex Incorporated · INSTRUMENT, BIOPSY3 reports
- ARROW IPN033803Teleflex Incorporated · INSTRUMENT, BIOPSY1 report
- ARROW IPN033811Teleflex Incorporated · INSTRUMENT, BIOPSY1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page