Elekta Medical Linear Accelerator recallZ-0116-2022 · Elekta, Inc. · initiated Sep 17, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.
FDA root cause category: Under Investigation by firm
What the company is doing
Written notification letter titled, "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" via email on 09/17/2021. The letter instructs consignees to please post the notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. Consignees are to advise the appropriate personnel, working with this product, on the content of the letter. The consignee is requested to submit an acknowledgement form to their local Elekta representative as soon as possible, but no longer than 30 days from receipt.
Recall details
- Recall number
- Z-0116-2022
- Status
- Terminated
- Initiated by the company
- Sep 17, 2021
- Recalling firm
- Elekta, Inc.
- Quantity
- 4,405 units
- Product code
- IYE
- FDA submissions
- K210500
- Distribution
- Distributed nationwide to FL, NC, OH, CA, IL, MD, OR, AL, NY, PA, MA, AK, GA, OR, KY, VA, TX, SD, AZ, NJ, MI, WI, ND, MO, WA, OK, PR, NM, NC, ID, CT, DE, LA, MT, TN, CO, UT, MN, IN, RI, IA, DC, ND, AZ, HI, KS, NE, NH, AR, WY, WV, MS, VT and internationally to Albania, Antigua/Barbuda, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kyrgystan, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe.
Product description
Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta Synergy; d. Elekta Infinity; e. Versa HD
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
a. Elekta Precise Digital Accelerator - UDI GTIN Number: 5060191071017, Model Numbers: XRT 0601 (NFB), XRT 0611 (FB), All Serial Numbers; b. Elekta Harmony - UDI GTIN Number: 05060191071581, Model Numbers: XRT 2201 (NFB), All Serial Numbers; c. Elekta Synergy - UDI GTIN Number: 05060191071550, Model Numbers: XRT 0621 (NFB), XRT 0631 (FB), All Serial Numbers; d. Elekta Infinity - UDI GTIN Number: 05060191071543, 05060191071512, Model Numbers: XRT 0511 (NFB & MLCi2), XRT 0521 (FB & MLCi2), XRT 0531 (NFB & Agility), XRT 0541 (FB & Agility), All Serial Numbers; e. Versa HD - UDI GTIN Number: 05060191071574, 05060191071529, Model Numbers: XRT 2121 (NFB), XRT 2131 (FB), All Serial Numbers
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- Elekta Medical Linear Accelerator 1552093Elekta Solutions AB · Accelerator, linear, medical5 recalls6 reports
- Elekta Medical Linear Accelerator 1552092Elekta Solutions AB · Accelerator, linear, medical5 recalls0 reports
- Elekta Medical Linear Accelerator 1550358Elekta Solutions AB · Accelerator, linear, medical4 recalls2 reports
- Elekta Medical Linear Accelerator 1552091Elekta Solutions AB · Accelerator, linear, medical4 recalls1 report
- Elekta Medical Linear Accelerator 1550359Elekta Solutions AB · Accelerator, linear, medical4 recalls0 reports
- Elekta Medical Linear Accelerator 1553032Elekta Solutions AB · Accelerator, linear, medical2 recalls0 reports
- NA 1525717Elekta Limited · Accelerator, linear, medical1 recall3 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page