Synchron recallZ-0399-2022 · Beckman Coulter Inc. · initiated Nov 10, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 11/10/21 recall notices were sent to customers. Customers with questions are encouraged to contact the Customer Support Center: http://www.beckmancoulter.com, 1-800-854-3633 in the United States and Canada. Customers were asked to take the following action: 1) Discontinue use of affected cholesterol reagent and follow the product replacement instructions. 2) Review the Safety Data Sheet (SDS) and discard according to your laboratory s Standard Operating Procedures and/or your local regulations. 3) Retrospective review is left to the discretion of the Laboratory Director. 4) Share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. 5) Complete and return the Response Form. For customers in the United States, if you need replacement product, complete the Replacement Order Form.

Recall details

Recall number
Z-0399-2022
Status
Open, Classified
Initiated by the company
Nov 10, 2021
Recalling firm
Beckman Coulter Inc.
Quantity
1433
Product code
CHH
FDA submissions
K971505
Distribution
US Nationwide distribution in the states of Louisiana, Ohio, North Carolina, California, South Carolina, Oklahoma, Pennsylvania, New York, Florida, New Jersey, Wisconsin, Alabama, Texas, Georgia, Maine, North Dakota, Virginia, Kansas, Washington, New Mexico, Iowa ,Nebraska, Idaho, Michigan, Maryland, Arizona, Hawaii, Missouri, Oregon, Mississippi, Rhode Island, Alaska, Tennessee, Illinois, Delaware, Kentucky, Arkansas, West Virginia, Massachusetts. OUS: India, South Africa, Saudi Arabia, Canada, Lebanon, Singapore, Philippines, Thailand

Product description

SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

Code information (models, lots, serial numbers)

Lot: M106450, UDI: 15099590575113

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.