Skip to content

Synchron 467825Recalls & reported problems

Beckman Coulter, Inc. · Cholesterol Reagent (CHOL 2 x 300) · Class 1

All models of Synchron (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Nov 10, 2021Class IIOpen, Classified

    Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.

    Z-0399-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code CHH (enzymatic esterase--oxidase, cholesterol) under 21 CFR 862.1175 as a Class I device. 21 CFR 862.1175 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Synchron 467825

Has the Synchron 467825 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0399-2022 (Class II), began on Nov 10, 2021: Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Synchron 467825 FDA approved?

Devices of this type (product code CHH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Beckman Coulter, Inc.

Who makes the Synchron 467825, and is it still sold?

The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand Synchron, model 467825). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Enzymatic Esterase--Oxidase, Cholesterol devices

All Enzymatic Esterase--Oxidase, Cholesterol devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page