Beckman Coulter, Inc. · 7 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one Synchron model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls
5

We found 7 models of Synchron by Beckman Coulter, Inc. in the FDA's device registry (GUDID). Together they are named in 4 adverse event reports in the FDA's MAUDE database, received between Mar 2, 2017 and Dec 19, 2018 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

5 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Synchron recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Sep 3, 2025Open, Classified

    Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.

    Z-0071-2026 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

  2. Nov 10, 2021Open, Classified

    Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.

    Z-0399-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  3. Sep 4, 2019Terminated

    Cross-reactivity testing may not meet the claims indicated in the Opiate 300 ng (OP)Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.

    Z-1752-2020 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

  4. Sep 4, 2019Terminated

    Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.

    Z-1751-2020 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

  5. Jun 12, 2019Terminated

    The firm has identified that cross-reactivity testing may not meet the claims indicated in the Barbiturates Reagent Chemistry Information Sheet (CIS). The presence of certain compounds in urine may result in false positive or false negative results.

    Z-1799-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported Synchron problems

Device problems

  • Low Test Results3
  • Incorrect Or Inadequate Test Results1

Patient problems

  • No Consequences Or Impact To Patient4

Synchron models

7 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Synchron adverse event reports

The 4 most recent reports across all models, in the reporter's own words as published by the FDA.