Skip to content

Medfusion Model 3500 and 4000 Syringe Infusion Pumps and… recallZ-0653-2022 · Smiths Medical ASD Inc. · initiated Dec 3, 2015

Class II Open, Classified All recalls posted in February 2022

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall occurred in 2015.

FDA root cause category: Component change control

What the company is doing

The firm issued Customer Information Bulletin dated 03-Dec-2015. It explained the details of the issue and provides the consignee a list of serial numbers for pumps in their facility that may be impacted. In addition, it informs the consignee if they have purchased main boards between April 2012 and June 2015 for part numbers G6000361, G6000435, G6001260, G6001560 and G6002728, they are included in the enclosed file. The consignee is to contact the recalling firm to make arrangements to begin the replacement process. If the Supercap Post alarm or the Backup Audible Alarm error code occurs prior to replacement, the consignee is to immediately remove the pump from service for repair by a trained biomedical technical at the recalling firm or at their facility. If the device with the affected serial number is sent to the recalling firm for any other unrelated repair, the supercapacitor will be replaced at that time. No response form was enclosed with the bulletin.

Recall details

Recall number
Z-0653-2022
Recall class
Class II
Status
Open, Classified
Initiated by the company
Dec 3, 2015
Posted by the FDA
Feb 17, 2022
Recalling firm
Smiths Medical ASD Inc.
Quantity
38,797 (31,065 pumps and 7,732 boards)
Product code
FRN
FDA submissions
K040899, K111386
Distribution
Distribution US nationwide, including PR, VI, GU, and MP. There was government/military/foreign distribution.

Product description

Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time period in which a main board was replaced using replacement main board part numbers G6000361, G6000435, G6001260, G6001560, G6001562 and G6002728.

Code information (models, lots, serial numbers)

Medfusion 3500 and 4000 Syringe Infusion Pumps manufactured between April 2012 and June 2015; and Medfusion 3010 and 3010A (and pump models 3500 and 40000 serviced during April 20121 and June 2015 in which a main board was replaced.

Other devices under the same FDA submission

15 devices cleared under K040899, K111386 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page