Medfusion Model 3500 and 4000 Syringe Infusion Pumps and… recallZ-0653-2022 · Smiths Medical ASD Inc. · initiated Dec 3, 2015
Class II Open, Classified All recalls posted in February 2022
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall occurred in 2015.
FDA root cause category: Component change control
What the company is doing
The firm issued Customer Information Bulletin dated 03-Dec-2015. It explained the details of the issue and provides the consignee a list of serial numbers for pumps in their facility that may be impacted. In addition, it informs the consignee if they have purchased main boards between April 2012 and June 2015 for part numbers G6000361, G6000435, G6001260, G6001560 and G6002728, they are included in the enclosed file. The consignee is to contact the recalling firm to make arrangements to begin the replacement process. If the Supercap Post alarm or the Backup Audible Alarm error code occurs prior to replacement, the consignee is to immediately remove the pump from service for repair by a trained biomedical technical at the recalling firm or at their facility. If the device with the affected serial number is sent to the recalling firm for any other unrelated repair, the supercapacitor will be replaced at that time. No response form was enclosed with the bulletin.
Recall details
- Recall number
- Z-0653-2022
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Dec 3, 2015
- Posted by the FDA
- Feb 17, 2022
- Recalling firm
- Smiths Medical ASD Inc.
- Quantity
- 38,797 (31,065 pumps and 7,732 boards)
- Product code
- FRN
- FDA submissions
- K040899, K111386
- Distribution
- Distribution US nationwide, including PR, VI, GU, and MP. There was government/military/foreign distribution.
Product description
Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time period in which a main board was replaced using replacement main board part numbers G6000361, G6000435, G6001260, G6001560, G6001562 and G6002728.
Code information (models, lots, serial numbers)
Medfusion 3500 and 4000 Syringe Infusion Pumps manufactured between April 2012 and June 2015; and Medfusion 3010 and 3010A (and pump models 3500 and 40000 serviced during April 20121 and June 2015 in which a main board was replaced.
Other devices under the same FDA submission
15 devices cleared under K040899, K111386 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Medfusion 4000-0106-01Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls4,390 reports
- Medfusion 4000-0105-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls2,882 reports
- Medfusion 3500-0600-01Icu Medical, Inc. · Pump, infusion2 recalls2,041 reports
- Medfusion 3500-500Icu Medical, Inc. · Pump, infusion2 recalls2 deaths1,884 reports
- Medfusion 4000-0101-51Icu Medical, Inc. · Pump, infusion2 recalls1 death1,795 reports
- Medfusion 3500-0600-82Icu Medical, Inc. · Pump, infusion2 recalls89 reports
- MEDFUSION 4000-0106-00Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls10 reports
- Medfusion 3500-0600-00Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls9 reports
- MEDFUSION 4000-0106-231Icu Medical, Inc. · Pump, infusion2 recalls5 reports
- Medfusion 3500-0600-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls2 reports
- Medfusion 4000-0105-249Icu Medical, Inc. · Pump, infusion2 recalls1 report
- Medfusion 4000-0105-78Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page