SOMATOM Confidence -Computed tomography systems recallZ-0472-2022 · Siemens Medical Solutions USA, Inc · initiated Nov 19, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis.

FDA root cause category: Software Design Change

What the company is doing

Siemens issued Urgent Medical Device Correction letter CSAN CT067/21/S on November 19, 2021, to notify affected customers of the head image degradation issue and to provide them with a workaround solution. Letter states reason for recall, health risk and action to take: To avoid potential misdiagnosis, it is mandatory to change to the new cupping correction setting CC = 4 immediately, restart the application, and (if applicable) reconstruct the raw data again. Cupping correction setting CC = 4 is a further development of CC = 1: The Image impression is very similar to setting CC = 1 but an important improvement reduces possible artifacts. For customers who have routinely been using CC = 0 and CC = 2 instead of CC = 1, the image impression with CC = 4 will be different from what they are used to. Please note that the use of cupping correction CC = 4 may cause slightly altered CT numbers of non-water equivalent body materials or tissues in reconstructions with H-kernels and enabled iBHC (i.e., setting Bone or Iodine ) compared to results obtained with settings CC = 0, 1, or 2. For any quantitative interpretation of CT numbers, e.g., the mapping of HU to electron density, a recalibration is necessary Siemens is developing a technical solution, VB20_SP6 with high priority .Siemens Healthineers is working with high priority on a solution for the described degradation. The solution will be provided with software version VB20_SP6. Your local service organization will contact you in due time and schedule an appropriate date for the installation of the solution. This solution will be provided to you free of charge. Please note that the Regional Unit Contact cannot schedule your appointment. If you have any unresolved questions or require technical support, please contact your local applicable specialist or our service organization at 1-800-888-7436. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the informati

Recall details

Recall number
Z-0472-2022
Status
Open, Classified
Initiated by the company
Nov 19, 2021
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
108 units
Product code
JAK
FDA submissions
K190578
Distribution
US Nationwide distribution.

Product description

SOMATOM Confidence -Computed tomography systems Model 10590100

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Serial Numbers: 100172 100091 100112 100362 100278 100158 100143 100129 100237 100238 100191 100356 100425 100216 100307 100325 100328 100329 100326 100239 100461 100213 100070 100062 100072 100022 100107 100241 100157 100055 100340 100347 100137 100267 100190 100252 100438 100046 100197 100092 100202 100313 100311 100043 100136 100189 100283 100156 100079 100049 100056 100147 100262 100018 100240 100117 100097 100274 100242 100099 100383 100059 100289 100312 100174 100073 100016 100182 100146 100058 100170 100175 100292 100090 100271 100463 100257 100273 100067 100160 100218 100367 100087 100309 100284 100207 100405 100261 100025 100338 100272 100053 100019 100421 100280 100357 100220 100214 100393 100394 100486 100140 100085 100279 100254 100353 100298 100379 UDI: UDI 04056869011011

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

5 devices cleared under K190578 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 89208).