SOMATOM Edge Plus-Computed tomography system recallZ-0474-2022 · Siemens Medical Solutions USA, Inc · initiated Nov 19, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis.
FDA root cause category: Software Design Change
What the company is doing
Siemens issued Urgent Medical Device Correction letter CSAN CT067/21/S on November 19, 2021, to notify affected customers of the head image degradation issue and to provide them with a workaround solution. Letter states reason for recall, health risk and action to take: To avoid potential misdiagnosis, it is mandatory to change to the new cupping correction setting CC = 4 immediately, restart the application, and (if applicable) reconstruct the raw data again. Cupping correction setting CC = 4 is a further development of CC = 1: The Image impression is very similar to setting CC = 1 but an important improvement reduces possible artifacts. For customers who have routinely been using CC = 0 and CC = 2 instead of CC = 1, the image impression with CC = 4 will be different from what they are used to. Please note that the use of cupping correction CC = 4 may cause slightly altered CT numbers of non-water equivalent body materials or tissues in reconstructions with H-kernels and enabled iBHC (i.e., setting Bone or Iodine ) compared to results obtained with settings CC = 0, 1, or 2. For any quantitative interpretation of CT numbers, e.g., the mapping of HU to electron density, a recalibration is necessary Siemens is developing a technical solution, VB20_SP6 with high priority .Siemens Healthineers is working with high priority on a solution for the described degradation. The solution will be provided with software version VB20_SP6. Your local service organization will contact you in due time and schedule an appropriate date for the installation of the solution. This solution will be provided to you free of charge. Please note that the Regional Unit Contact cannot schedule your appointment. If you have any unresolved questions or require technical support, please contact your local applicable specialist or our service organization at 1-800-888-7436. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the informati
Recall details
- Recall number
- Z-0474-2022
- Status
- Open, Classified
- Initiated by the company
- Nov 19, 2021
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 80 units
- Product code
- JAK
- FDA submissions
- K190578
- Distribution
- US Nationwide distribution.
Product description
SOMATOM Edge Plus-Computed tomography system Model 1026700
Code information (models, lots, serial numbers)
Serial Numbers: 122022 122185 122098 122062 122099 122121 122146 122054 122082 122080 122020 122164 122116 122008 122036 122029 122044 122172 122074 122075 122049 122194 122141 122167 122156 122183 122081 122126 122129 122090 122130 122179 122089 122085 122037 122009 122163 122177 122066 122064 122038 122161 122132 122171 122153 122168 122158 122144 122063 122086 122035 122041 122112 122114 122107 122101 122108 122109 122162 122187 122174 122125 122178 122011 122182 122052 122034 122057 122145 122147 122050 122106 122189 122180 122061 122051 122166 122170 122092 122004 UDI: UDI 04056869006949
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- System SOMATOM Definition Edge 10590000Siemens Healthcare GmbH · System, x-ray, tomography, computed2 recalls1 death11 reports
Other devices under the same FDA submission
5 devices cleared under K190578 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- System SOMATOM Definition AS 08098027Siemens Healthcare GmbH · System, x-ray, tomography, computed1 recall1 death47 reports
- System SOMATOM Confidence 10590100Siemens Healthcare GmbH · System, x-ray, tomography, computed1 recall2 reports
- Somatom Force 10742326Siemens Healthcare GmbH · System, x-ray, tomography, computed3 deaths28 reports
- SOMATOM Definition Flash 10430603Siemens Healthcare GmbH · System, x-ray, tomography, computed1 death27 reports
- System SOMATOM Drive 10431700Siemens Healthcare GmbH · System, x-ray, tomography, computed2 deaths15 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89208).
- Z-0472-2022 · Open, ClassifiedSOMATOM Confidence -Computed tomography systems Model 10590100
- Z-0473-2022 · Open, ClassifiedSOMATOM Definition AS-Computed tomography system Model 8098027
- Z-0475-2022 · Open, ClassifiedSOMATOM Definition Edge -Computed tomography system Model 10590000
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page