Medical device recalls: May 2022
The FDA posted 13 medical device recall records in May 2022, from 11 recall events by 11 companies. 7 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Atrium Medical Corporation · 1 event, 2 records, 2 Class I
- Baxter Healthcare Corporation · 1 event, 2 records, 2 Class I
- Abbott Vascular · 1 event, 1 record, 1 Class I
- ARJOHUNTLEIGH POLSKA Sp. z.o.o. · 1 event, 1 record, 1 Class I
- Avanos Medical, Inc. · 1 event, 1 record, 1 Class I
- Ion Beam Applications S.A. · 1 event, 1 record
- Luminex Corporation · 1 event, 1 record
- Medtronic Neuromodulation · 1 event, 1 record
FDA root cause
- Device Design · 8
- Under Investigation by firm · 4
- Software design · 1
Class I recalls (7)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- In-Line ventilator adaptorClass IZ-1138-2022 · Baxter Healthcare Corporation · posted May 30, 2022 · Terminated
There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.
- In-Line ventilator adaptorClass IZ-1139-2022 · Baxter Healthcare Corporation · posted May 30, 2022 · Terminated
There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.
- Z-1098-2022 · ARJOHUNTLEIGH POLSKA Sp. z.o.o. · posted May 21, 2022 · Terminated
The device may emit smoke or ignite.
- Dragonfly OpStar Imaging Catheter, REF 1014651/1014652, 0.068 in, 0.014 in, 135 cm, STERILE EOClass IZ-1095-2022 · Abbott Vascular · posted May 19, 2022 · Open, Classified
There is a potential that the proximal marker may separate from the imaging catheter
- Z-1076-2022 · Atrium Medical Corporation · posted May 18, 2022 · Open, Classified
Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible administration of additional anesthesia and…
- iCast Covered Stent SystemClass IZ-1077-2022 · Atrium Medical Corporation · posted May 18, 2022 · Open, Classified
Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible administration of additional anesthesia and…
- Z-1032-2022 · Avanos Medical, Inc. · posted May 13, 2022 · Terminated
Product labeling is being modified. STOP using the Anonymous Account Mode feature of the CORTRAK 2 device immediately. CONFIRM placement of the NG/NI tube per institution protocol.
Class II recalls (6)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1205-2022 · Ion Beam Applications S.A. · posted May 27, 2022 · Open, Classified
First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed
- Z-1123-2022 · Medtronic Neuromodulation · posted May 23, 2022 · Open, Classified
The Implantable Neurostimulator Model may be susceptible to not communicating with the clinical programmer and/or patient programmer system.
- Z-1111-2022 · Turbett Surgical, LLC · posted May 18, 2022 · Open, Classified
Eight units manufactured before June 2019 have exhibited weld failures breaching the sterile barrier after prolonged use.
- Z-1092-2022 · Luminex Corporation · posted May 13, 2022 · Open, Classified
Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.
- Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging.Class IIZ-1034-2022 · Olympus Corporation of the Americas · posted May 5, 2022 · Open, Classified
GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients
- Z-0886-2022 · Philips Medical Systems Nederland B.V. · posted May 3, 2022 · Open, Classified
In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.