Volara System recallZ-1138-2022 · Baxter Healthcare Corporation · initiated Apr 26, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.

FDA root cause category: Device Design

What the company is doing

The recall notification was sent to consignees on 04/26/2022 via FedEx. The customer was instructed to sign acknowledgement form to confirm receipt and return to Hillrom using pre-paid envelope.

Recall details

Recall number
Z-1138-2022
Status
Terminated
Initiated by the company
Apr 26, 2022
Posted by the FDA
May 30, 2022
Recalling firm
Baxter Healthcare Corporation
Quantity
259 units
Product code
NHJ
FDA submissions
K200988
Distribution
US Nationwide distribution in the states of TX, FL, VA, KY, CA, AZ, WI, IL, NC, OH, MI, MS, MO, MN, AL, TN, SD, OK, NV, NJ, IN, CO, GA, WA, PA, IA, NE, NM, SC, ND, KS, WV, OR.

Product description

In-Line ventilator adaptor

Code information (models, lots, serial numbers)

M08473 OPTIMUS OLE AC PAT.CIRCUIT KIT; UDI-DI: 00887761981492. Manufacturing dates February 28, 2020 to present.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 90061).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page