Skip to content

MicroScan recallZ-1085-2022 · Beckman Coulter Inc. · initiated Oct 21, 2019

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

FDA root cause category: Mixed-up of materials/components

What the company is doing

During the week of 10/21/2019, Beckman Coulter sent an "URGENT MEDICAL DEVICE RECALL" Letter via e-mail and postal mail informing customers that some WalkAway instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1A (ampere) current rating rather than the correct current rating at 10A. Customers are instructed to: If the component fails and and the WalkAway suffers a power loss, user should: --Contact the Customer Support Center. If in the United States call 1-800-677-7226. If outside the United States, contact their local Beckman Coulter Representive. --Determine if the customer knows how long the WalkAway has been without power. If the time is unknown or greater than 2 hours, then panels should be reset, or an alternate test method should be used. -- If within 2 hours window the instrument will maintain the environmental conditions. Depending on the panel type, the user has the option to read the panels manually, on an autoSCAN-4 (if available) or reset the panels. - If the component remains functional, a Beckman Coulter Field Service Representative will contact the user to schedule an onsite visit. During the visit the representative will inspect the suspect instrument and if the incorrect component is found, will replace the component with the correct one rated at 10A. -Complete and return Customer Response Form to Beckman Coulter, Inc. 1584 Enterprise Blvd. West Sacramento, CA 95691 Attn: Quality Systems & Compliance Fax number: 916-374-2119 Email address: [email protected] For questions, contact: -via website, http://www.beckmancoulter.com/customersupport/support -via phone, call 1-800-677-7226 in the United States, -If outside the United States, contact the local Beckman Coulter Representative.

Recall details

Recall number
Z-1085-2022
Status
Open, Classified
Initiated by the company
Oct 21, 2019
Recalling firm
Beckman Coulter Inc.
Quantity
40 systems
Product code
LRG
FDA submissions
K911400
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The countries of Belgium, Brazil, Colombia, Germany, Hong Kong, Indonesia, Italy, Japan, Paraguay, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, United Arab Emirates, Vietnam and Yemen.

Product description

DxM 1040 MicroScan WalkAway Instrument REF B1018-440

Code information (models, lots, serial numbers)

Catalog Number: B1018-440 UDI-DI Code: 15099590690885 Serial Numbers: 44010026 44010027 44010028 44010029 44010030 44010031 44010032 44010033 44010035 44010036 44010037 44010038 44010040 44010041 44010042 44010043 44010044 44010045 44010046 44010047 44010048 44010049 44010050 44010051 44010052 44010053 44010054 44010055 44010056 44010057 44010058 44010059 44010060 44010061 44010063 44010064 44010065 44010066 44010067 44010068

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 90087).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page