Skip to content

MicroScan B1018-440Recalls & reported problems

Beckman Coulter, Inc. · DxM 1040 MicroScan WalkAway Instrument · Class 2

All models of MicroScan (10)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
1
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15099590690885), received between Jun 2, 2020 and Feb 13, 2025 — classified as 1 death, 1 malfunction.

The most frequently reported specific device problem is “Device Misassembled During Manufacturing /Shipping” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 21, 2019Class IIOpen, Classified

    WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

    Z-1085-2022 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Misassembled During Manufacturing /Shipping1
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1
  • Not Applicable1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED GETTING HIGH PROBABILITY MISIDENTIFICATION (HPID) OF CITROBACTER FARMERI ON PATIENT SAMPLES USING NEG URINE COMBO 90 (NUC90, P/N C48640, LOT # 2025-11-11) PANELS PROCESSED IN THEIR DXM 1040 MICROSCAN WALKAWAY INSTRUMENT S/N (B)(6). THE CUSTOMER STATED THE MISIDENTIFICATION IMPACTED ONE PATIENT SAMPLE, MULTIPLE CULTURES. THE CUSTOMER NOTIFIED THE PHYSICIAN OF THE ERRONEOUS RESULTS OBTAINED FROM URINE SAMPLE. THE CUSTOMER STATED THE PATIENT WAS PUT ON HOSPICE AND PASSED AWAY ON (B)(6) 2025. IT WAS FURTHER NOTED THAT THE PATIENT DID NOT HAVE CAPACITY AND WAS REFUSING TREATMENT. THE PATIENT¿S FAMILY OPTED FOR COMFORT FOCUSED MEASURES. THE CUSTOMER INDICATED THEY REACHED OUT TO THEIR INTERNAL TEAM, AND THEY STATED THE MISIDENTIFICATION CAUSED A DELAY IN APPROPRIATE TREATMENT FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21381586 ↗
  • MalfunctionJun 2, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) REPORTED THAT DURING THE UNPACKING OF THE DXM 1040 WALKAWAY INSTRUMENT, THE COUNTERWEIGHT, COUNTER WEIGHT ROD AND VERTICAL AXIS READ ASSEMBLY SLIPPED CAUSING A NEAR INJURY TO THE FSE DUE TO A MISSING SHAFT LOCK CLIP. THE FSE WAS NOT INJURED AS THE PACKAGING MATERIAL CAUGHT THE COUNTERWEIGHT AFTER ONLY 1-INCH SLIP.

    Patient problems: No Consequences Or Impact To Patient; No Patient Involvement; Not Applicable

    MAUDE report 10111548 ↗

FDA clearance and classification

The FDA regulates product code LRG (instrument for auto reader & interpretation of overnight suscept. systems) under 21 CFR 866.1640 as a Class II device. 21 CFR 866.1640 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the MicroScan B1018-440

Has the MicroScan B1018-440 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1085-2022 (Class II), began on Oct 21, 2019: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the MicroScan B1018-440?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jun 2, 2020 and Feb 13, 2025. In those reports, the most common specific device problems are “Device Misassembled During Manufacturing /Shipping” (1) and “Incorrect, Inadequate or Imprecise Result or Readings” (1); the most common patient problem is “Not Applicable” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the MicroScan B1018-440?

1 report is classified as a death, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the MicroScan B1018-440, and is it still sold?

The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand MicroScan, model B1018-440). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems devices

All Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page