Skip to content

NIM Vital recallZ-1224-2022 · Medtronic Xomed, Inc. · initiated May 3, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Xomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software anomaly was identified.

FDA root cause category: Software Design Change

What the company is doing

The firm issued a recall notification to consignees on 05/03/2022 via mail. The letter instructs the consignee to follow the IFU regarding the setup and "Lead Off Detected" error message, forward the notification to those who need to be aware within the organization and to any location which the devices have been distributed, and to complete the response form. An additional correction notification was sent on about 09/13/2023 to notify consignees of the update availability (NIM Vital" SW version v1.4.3) to fix this issue.

Recall details

Recall number
Z-1224-2022
Status
Open, Classified
Initiated by the company
May 3, 2022
Posted by the FDA
Jun 2, 2022
Recalling firm
Medtronic Xomed, Inc.
Quantity
29 units
Product code
GWF
FDA submissions
K200759
Distribution
Worldwide distribution - US Nationwide distribution in the states of TX, CO, NJ and the country of France.

Product description

NIM Vital Patient Interface 4.0 (P/N: NIM4CPB1) NIM with Software version 1.3.2

Code information (models, lots, serial numbers)

Model Number: NIM4CPB1; UDI/DI: 00763000002985, lot numbers: P1910256, P1910260, P1910402, P1910404, P2008073, P2008130, P2008131 and 00763000395902, lot numbers: P2026275, P2026290, P2026329, P2026330, P2026357, P2026388, P2026389, P2026680, P2026681, P2026718, P2026719, P2026793, P2026956, P2026957, P2026958, P2026959, P2026981, P2026994, P2026995, P2127005, P2127006, P2127013

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

7 devices cleared under K200759 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 90163).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page