CARESCAPE Central Station recallZ-1265-2022 · GE Healthcare, LLC · initiated May 10, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.
FDA root cause category: Process control
What the company is doing
The recalling firm issued letters flagged URGENT MEDICAL DEVICE CORRECTION dated 5/10/2022 via traceable means. The letter listed the affected product details and described the safety issue and actions to be taken by the customer/user. It said they can continue to use the CSCS V2 to monitor patients on the CARESCAPE Network. In the event there is a loss of patient monitoring at the CSCS V2, they were to follow the safety instructions, which included: Ensure all patients are adequately monitored using other GE Central Stations, with alternate monitoring devices, or with increased patient surveillance. If additional GE Healthcare Clinical Information Center (CIC) or CARESCAPE Central Station devices are available in the same care unit, unmonitored patients will be automatically assigned to any open, unlocked Multi-Viewer patient windows. If there are not enough Multi-Viewer patient windows, other GE Central Stations in the same care unit will provide a list of unmonitored patients at the top of the Multi-Viewer display. Unmonitored patients can also be viewed on and assigned to any GE Central Station connected to the CARESCAPE Network in a different care unit, and the methods for doing this were described. The customer is referred to the Managing Patients chapter of their Clinical Information Center (CIC) or CARESCAPE Central Station (CSCS) User Manual for more information regarding patient viewing. The letter informed the customer that GE Healthcare will correct all affected products when the correction is available. A Medical Device Notification Acknowledgement Response Required form was enclosed to be completed and returned within 30 days from receipt by scanning or taking a photo of the completed form and emailing it back to the firm. The form is to confirm receipt and understanding of the Correction Notice. UPDATE: On 2/6/2023, the recalling firm issued an amended recall letter dated 2/6/2023 informing the consignee they were now including certain Fi
Recall details
- Recall number
- Z-1265-2022
- Status
- Open, Classified
- Initiated by the company
- May 10, 2022
- Posted by the FDA
- Jun 13, 2022
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 30,993 Central Stations and 2,180 FRUs and displays
- Product code
- DXJ
- FDA submissions
- K162012
- Distribution
- Worldwide distribution - US Nationwide. There was also military/government distribution and the countries of Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Kazakhstan, Kenya, Korea, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Macedonia, Madagascar, Malaysia, Maldives, Malta, Mauritius, Mexico, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, and Yemen.
Product description
CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part numbers have been added as affected devices: (1) Part #2082294-001, FRU MP200 POWER SUPPLY ASSEMBLY; (2) Part #2082293-002-R, FRU MP200 CPU ASSEMBLY; (3) Part #2082293-002, FRU MP200 CPU ASSEMBLY; (4) Part #2071523-001, MP200 22 IN DISPLAY WITH TOUCH; (5) Part #2092930-001, DSP LCD 22 IN MED W/TCH W/DKST N AMR; (6) Part #2092930-002, DSP LCD 22 IN MED W/TCH W/DKST BRIT P/C; (7) Part #2092930-003, DSP LCD 22IN MED W/TCH W/DKST EURO P/C; (8) Part #2092930-004, DSP LCD 22IN MED W/TCH W/DKST ITA P/C; (9), Part #2092930-005, DSP LCD 22IN MED W/TCH W/DKST INDIAN P/C; (10) Part #2092930-006, DSP LCD 22IN MED W/TCH W/DKST ISREAL P/C; (11) Part #2092930-007, DSP LCD 22IN MED W/TCH W/DKST SWISS P/C; (12) Part #2092930-008, DSP LCD 22IN MED W/TCH W/DKST AUSTRL P/C; (13) Part #2092930-009, DSP LCD 22IN MED W/TCH W/DKST JPN; (14) Part #2092930-010, DSP LCD 22IN MED W/TCH W/DKST BRZL P/C; (15) Part #2092930-011, DSP LCD 22IN MED W/TCH W DKST ARGNT P C; (16) Part #2092930-012, DSP LCD 22IN MED W TCH W DKST DENMARK; (17) Part #2092930-013, DSP LCD22IN MED W TCH W DKST S AFRICA P C; and (18) Part #2092930-014, DSP LCD22IN MED W TCH W DKST CHINA P C.
Code information (models, lots, serial numbers)
All CARESCAPE Central Station V2 units: (1) CSCS V2 MA1700 Integrated, Product Code SKN, Model #2082278-001, GTIN 00840682109666; (2) CSCS V2 MAS700 Desktop, Product Code SNF, Model #2082279-001, GTIN 00840682109604. NOTE: The Product Code is the first 3 letters of the serial number.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- CARESCAPE Central Station V2 Standard MAS 700Ge Medical Systems Information Technologies, Inc. · DETECTOR AND ALARM, ARRHYTHMIA3 recalls0 reports
- CARESCAPE Central Station V2 Integrated MAI 700Ge Medical Systems Information Technologies, Inc. · DETECTOR AND ALARM, ARRHYTHMIA3 recalls0 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K162012 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CARESCAPE CENTRAL STATION MAI800Ge Medical Systems Information Technologies, Inc. · Computer, Diagnostic, Programmable3 recalls0 reports
- CARESCAPE CENTRAL STATION MAI800Ge Medical Systems Information Technologies, Inc. · Computer, Diagnostic, Programmable2 recalls1 death1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page