Skip to content

CARESCAPE Central StationRecalls & reported problems across all models

Ge Medical Systems Information Technologies, Inc. · 4 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

5 open recalls name at least one CARESCAPE Central Station model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls
6

We found 4 models of CARESCAPE Central Station by Ge Medical Systems Information Technologies, Inc. in the FDA's device registry (GUDID). Together they are named in 1 adverse event report in the FDA's MAUDE database, received between Feb 2, 2026 and Feb 2, 2026 — classified as 1 death.

The most frequently reported specific device problem is “No Audible Alarm” (1 report).

6 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

CARESCAPE Central Station recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jul 30, 2026Open, Classified

    Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

    Z-3143-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  2. Sep 5, 2025Open, Classified

    Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

    Z-0214-2026 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  3. Sep 5, 2025Open, Classified

    Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

    Z-0213-2026 · Root cause: Software design · Recall details · FDA record ↗

  4. May 10, 2022Open, Classified

    The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.

    Z-1265-2022 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗

  5. Jan 28, 2022Open, Classified

    If the CARESCAPE Central Station v2.0 is used with an unapproved keyboard, the audio can be muted resulting in loss of audible alarms.

    Z-0808-2022 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  6. Jan 11, 2018Terminated

    CARESCAPE Central Station (CSCS) software version 2.0.2 units may experience unexpected NO COMM (No Communication) and network communication issues after boot-up or system restart.

    Z-0487-2018 · Root cause: Software in the Use Environment · names 2 of these models · Recall details · FDA record ↗

Most reported CARESCAPE Central Station problems

Device problems

  • No Audible Alarm1

Patient problems

  • Insufficient Information1

CARESCAPE Central Station models

4 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent CARESCAPE Central Station adverse event reports

The 1 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Ge Medical Systems Information Technologies, Inc. products

All Ge Medical Systems Information Technologies, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page