CARESCAPE CENTRAL STATION MAI800 (REF 5867474-03)Recalls & reported problems
Ge Medical Systems Information Technologies, Inc. · MAI800 MEDICAL DEVICE · Class 2
All models of CARESCAPE Central Station (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00195278343383), received between Feb 2, 2026 and Feb 2, 2026 — classified as 1 death.
The most frequently reported specific device problem is “No Audible Alarm” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (K162012); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 30, 2026Class IIOpen, Classified
Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
Z-3143-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Sep 5, 2025Class IIOpen, Classified
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
Z-0214-2026 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- No Audible Alarm1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathFeb 2, 2026No Audible Alarm
IT WAS REPORTED THAT A PATIENT WAS FOUND UNRESPONSIVE AND THE CENTRAL STATION MAY NOT HAVE AUDIBLY ALARMED. THE PATIENT WAS TRANSFERRED TO THE ICU WHERE THEY EVENTUALLY EXPIRED.
Patient problems: Insufficient Information
MAUDE report 24240085 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K162012. One submission often covers many models, so check the product and lot details in each record.
- Z-0213-2026 · Sep 5, 2025 · Open, ClassifiedGE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube
- Z-1265-2022 · May 10, 2022 · Open, ClassifiedCARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part numbers have been added as affected devices:
- Z-0808-2022 · Jan 28, 2022 · Open, ClassifiedGE Healthcare CARESCAPE Central Station (CSCS) V2.0x MAI700 (Integrated), REF 2082278-001, and MAS700 (Desktop, REF 2082279-001.
- Z-0487-2018 · Jan 11, 2018 · TerminatedGE HEALTHCARE CARESCAPE Central Station (CSCS) software version 2.0.2 The CARESCAPE Central Station is intended to collect information from a network and display this data. This
Other Computer, Diagnostic, Programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page