CARESCAPE CENTRAL STATION MAI800 (REF 5867474-03)Recalls & reported problems

Ge Medical Systems Information Technologies, Inc. · MAI800 MEDICAL DEVICE · Class 2

All models of CARESCAPE Central Station (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00195278343383), received between Feb 2, 2026 and Feb 2, 2026 — classified as 1 death.

The most frequently reported specific device problem is “No Audible Alarm” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

4 other recalls cover devices cleared under the same FDA submission (K162012); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Class IIOpen, Classified

    Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

    Z-3143-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Sep 5, 2025Class IIOpen, Classified

    Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

    Z-0214-2026 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Audible Alarm1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 2, 2026No Audible Alarm

    IT WAS REPORTED THAT A PATIENT WAS FOUND UNRESPONSIVE AND THE CENTRAL STATION MAY NOT HAVE AUDIBLY ALARMED. THE PATIENT WAS TRANSFERRED TO THE ICU WHERE THEY EVENTUALLY EXPIRED.

    Patient problems: Insufficient Information

    MAUDE report 24240085 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K162012. One submission often covers many models, so check the product and lot details in each record.

  • Z-0213-2026 · Sep 5, 2025 · Open, Classified
    GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube
  • Z-1265-2022 · May 10, 2022 · Open, Classified
    CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part numbers have been added as affected devices:
  • Z-0808-2022 · Jan 28, 2022 · Open, Classified
    GE Healthcare CARESCAPE Central Station (CSCS) V2.0x MAI700 (Integrated), REF 2082278-001, and MAS700 (Desktop, REF 2082279-001.
  • Z-0487-2018 · Jan 11, 2018 · Terminated
    GE HEALTHCARE CARESCAPE Central Station (CSCS) software version 2.0.2 The CARESCAPE Central Station is intended to collect information from a network and display this data. This

Other Computer, Diagnostic, Programmable devices

All Computer, Diagnostic, Programmable devices →