Artis zeego recallZ-1343-2022 · Siemens Medical Solutions USA, Inc · initiated May 24, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.

FDA root cause category: Component design/selection

What the company is doing

On May 24, 2022, the firm mailed Urgent Medical Device Correction letters to affected consignees. Customers were informed that if the system is not used regularly, please ensure that the system is switched on in due time before any use to check the status. Please ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. The firm will replace the Bios battery and the service interval will be shortened to two years. The firm will contact customers to arrange a date to perform the correction. Customers may call the service organization at 1-800-888-7435 to schedule an earlier appointment.

Recall details

Recall number
Z-1343-2022
Status
Terminated
Initiated by the company
May 24, 2022
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
233 worldwide; 32 US
Product code
OWB
FDA submissions
K181407
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, FL, IL, LA, ME, MN, MO, NC, OH, PA, SC, TN, TX, WI and the countries of Australia, Austria, Azerbaijan, Belgium, Canada, China, Finland, France, Germany, Hong Kong, India, Italy, Japan, Lebanon, Lithuania, Luxembourg, Netherlands, Poland, Portugal, South Korea, Switzerland, United Kingdom.

Product description

Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System

Code information (models, lots, serial numbers)

UDI-DI: Artis zeego 04056869010083 Artis Q.zeego 04056869010007

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

10 devices cleared under K181407 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page