Olympus recallZ-1814-2022 · Olympus Corporation of the Americas · initiated Aug 12, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
FDA root cause category: Under Investigation by firm
What the company is doing
Olympus issued Urgent Medical Device Recall Letter on 8/12/22 to Surgical Department Risk Management. Letter states reason for recall, health risk and action to take: 1. Immediately locate any product with affected lot numbers listed in this communication. See Attachment 2 for pictures showing the location of lot numbers on the package. 2. Call your Olympus customer service representative at 1 800 848 9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number 0412 and provide your contact information as indicated in the portal. 4. If you have distributed these devices outside your facility, please notify your customers of this removal action immediately. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Contact directly at (647) 999-3203 or at [email protected] for any additional information or support concerning this matter.
Recall details
- Recall number
- Z-1814-2022
- Status
- Open, Classified
- Initiated by the company
- Aug 12, 2022
- Posted by the FDA
- Sep 26, 2022
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 231.4 units (5/box); 510 units (5/box) OUS
- Product code
- GEI
- FDA submissions
- K212643
- Distribution
- Nationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREA
Product description
Olympus POWERSEAL 5MM, 37CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0537CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surgical procedures.
Code information (models, lots, serial numbers)
UDI-DI: 00821925044555 Lot Numbers: CA191233 CA191234 CA191235 CA233514 CA233515 CA233516 CA233517 CA233518 CA236206 CA236208 CA236213 CA236215 CA236216 CA236217 CA236220 CA236227 CA236228 CA236230 CA236238 CA236240 CA236243 CA236261 CA236272 CA236273 CA236276 CA236278 CA236290
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Olympus PS-0537CJDAGyrus ACMI, LLC · Electrosurgical, cutting & coagulation & accessories2 recalls184 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K212643 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Olympus PS-0523CJDAGyrus ACMI, LLC · Electrosurgical, cutting & coagulation & accessories3 recalls90 reports
- Olympus PS-0544CJDAGyrus ACMI, LLC · Electrosurgical, cutting & coagulation & accessories2 recalls10 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90788).
- Z-1813-2022 · Open, ClassifiedOlympus POWERSEAL 5MM, 23CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0523CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surg…
- Z-1815-2022 · Open, ClassifiedOlympus POWERSEAL 5MM, 44CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0544CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surg…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.