Skip to content

Olympus PS-0537CJDARecalls & reported problems

Gyrus ACMI, LLC · POWERSEAL 5 MM, 37 CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION (5/PK) · Class 2

All models of Olympus (27)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus PowerSeal sealer has 174 adverse event reports, with no deaths and 14 injuries. The reports mostly describe tissue that fails to seal, the device not activating or delivering energy, and parts breaking. Two recalls name this device, both open: one covers an error that stops energy delivery, and the other covers a loose rotation collar that can delay treatment. Reports rose to 53 in the latest 12 months from 19 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
184
Deaths reported
0
Injuries reported
14
Recalls naming it
2

Summary

The FDA's MAUDE database contains 184 adverse event reports that list this exact device (DI 00821925044555), received between Jul 25, 2022 and Sep 30, 2026 — classified as 14 injuries, 170 malfunctions.

The most frequently reported specific device problem is “Failure to Seal” (52 reports).

2 FDA recall records name this device's identifier, 2 still open.

5 other recalls cover devices cleared under the same FDA submission (K212643); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 4 reportsMar 2024: 2 reportsApr 2024: 8 reportsMay 2024: 6 reportsJun 2024: 8 reportsJul 2024: 22 reportsAug 2024: 9 reportsSep 2024: 12 reportsOct 2024: 8 reportsNov 2024: 6 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 14 reportsJan 2026: 5 reportsFeb 2026: 7 reportsMar 2026: 4 reportsApr 2026: 9 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 4 reportsSep 2026: 13 reportsOct 2026: 0 reports20242025202622
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 7, 2024Open, Classified

    Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.

    Z-1700-2024 · Root cause: Process control · Recall details · FDA record ↗

  2. Aug 12, 2022Open, Classified

    An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment

    Z-1814-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Seal52
  • Activation Failure17
  • Break13
  • Communication or Transmission Problem10
  • Energy Output Problem10
  • Failure to Deliver Energy8
  • Activation Problem6
  • Detachment of Device or Device Component6

Patient problems

  • No Clinical Signs, Symptoms or Conditions177
  • Hemorrhage/Blood Loss/Bleeding4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Foreign Body In Patient1
  • Loss of consciousness1
  • Paralysis1

Recent adverse event reports

The 25 most recent of 184 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Contamination /Decontamination Problem

    The customer reported that there was a black stain adhered to the handle section involving an Olympus Powerseal. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948840 ↗
  • MalfunctionSep 28, 2026Break

    The customer reported tip breakage during a therapeutic tumorectomy procedure involving a PowerSeal. There was no delay and the procedure was complete with an Olympus back up device. There was no patient harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50913316 ↗
  • MalfunctionSep 25, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902637 ↗
  • MalfunctionSep 24, 2026Contamination /Decontamination Problem

    The customer reported that there were black stains at multiple locations on the main body, involving an Olympus Power Seal curved jaw sealer. There was no procedure involved. There was no patient injury or harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883214 ↗
  • MalfunctionSep 24, 2026Contamination /Decontamination Problem

    The customer reported that there were black stains at multiple locations on the main body, involving an Olympus Power Seal curved jaw sealer. There was no procedure involved. There was no patient injury or harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883032 ↗
  • MalfunctionSep 24, 2026Contamination /Decontamination Problem

    The customer reported that there were black stains at multiple locations on the main body, involving an Olympus Power Seal curved jaw sealer. There was no procedure involved. There was no patient injury or harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882923 ↗
  • MalfunctionSep 21, 2026Failure to Power Up

    The customer reported that the subject device did not power on the first attempt, and even after trying about three times, there was no change in function, involving an Olympus Zoom Microscope Body. The issue was found during a Left adrenalectomy Therapeutic procedure. The intended procedure was completed with an Olympus back-up device. There was no patient injury or harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50845930 ↗
  • MalfunctionSep 16, 2026Failure to Seal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50796444 ↗
  • MalfunctionSep 16, 2026Failure to Seal

    The customer reported that the subject device had poor grounding, involving an Olympus PowerSeal curved Jaw sealer. The issue was found during an unknown Therapeutic procedure. There was no patient injury or harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50793821 ↗
  • MalfunctionSep 15, 2026Failure to Seal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50779634 ↗
  • MalfunctionSep 14, 2026Failure to Deliver Energy; Key or Button Unresponsive/not Working

    The customer reported that after turning off the power once and trying to use POWERSEAL again, an i0486 error occurred and it became unusable, involving an Olympus PowerSeal curved Jaw sealer. The issue was found during a left hemicolectomy Therapeutic procedure. The intended procedure was completed with an Olympus back-up device. There was no patient injury or harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755366 ↗
  • MalfunctionSep 10, 2026Connection Problem; Activation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50724960 ↗
  • MalfunctionSep 10, 2026Failure to Seal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719230 ↗
  • MalfunctionAug 31, 2026Activation Failure; Physical Resistance/Sticking

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50519013 ↗
  • MalfunctionAug 14, 2026Failure to Seal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50254545 ↗
  • MalfunctionAug 13, 2026Failure to Cycle

    No additional information received from customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50243891 ↗
  • MalfunctionAug 13, 2026Failure to Seal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50239178 ↗
  • MalfunctionJul 31, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50098867 ↗
  • MalfunctionJul 30, 2026No Device Output

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50083059 ↗
  • MalfunctionJul 27, 2026Failure to Seal

    The customer reported seal not completed appeared and sealing stopped during a therapeutic thoracoscopic esophagectomy procedure involving an Olympus Powerseal. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50019826 ↗
  • MalfunctionMay 19, 2026Break

    IT WAS REPORTED THE POWERSEAL'S JAW BROKE DURING THE UBERECTOMY THERAPEUTIC PROCEDURE. THE PROCEDURE WAS DELAYED BY FOUR MINUTES DURING SEDATION WHEN THE DEVICE WAS OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACKUP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25218021 ↗
  • MalfunctionMay 6, 2026Failure to Seal

    IT WAS REPORTED THE POWERSEAL HAD INCOMPLETE SEAL CYCLE ERROR. THE ISSUE WAS FOUND DURING HYSTERECTOMY, THERAPEUTIC PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT INJURY OR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25094514 ↗
  • MalfunctionMay 5, 2026Activation Failure

    IT WAS REPORTED THAT THE POWERSEAL FAILED TO SEAL. THE ISSUE WAS FOUND DURING A THERAPEUTIC LAPAROSCOPIC HYSTERECTOMY AND WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25087790 ↗
  • MalfunctionMay 1, 2026Device Damaged Prior to Use

    IT WAS REPORTED THE POWERSEAL WOULD NOT POWER UP WHEN CONNECTED TO THE GENERATOR. THE ISSUE WAS FOUND DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25061983 ↗
  • MalfunctionApr 27, 2026Retraction Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE POWERSEAL BLADE WAS STUCK AND COULD NOT BE CLOSED. THE ISSUE WAS FOUND DURING AN UNKNOWN PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25010393 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K212643. One submission often covers many models, so check the product and lot details in each record.

  • Z-3035-2026 · Aug 18, 2026 · Open, Classified
    POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories
  • Z-1701-2024 · Mar 7, 2024 · Open, Classified
    POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascul
  • Z-1699-2024 · Mar 7, 2024 · Open, Classified
    POWERSEAL 5 mm, 23 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascul
  • Z-1815-2022 · Aug 12, 2022 · Open, Classified
    Olympus POWERSEAL 5MM, 44CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0544CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surg
  • Z-1813-2022 · Aug 12, 2022 · Open, Classified
    Olympus POWERSEAL 5MM, 23CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0523CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surg

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page